9

Using the Front Panel Display with the Power Analyzer Input

To select the measurement source:
Action Display
Press Meter to return the display to Meter mode. Press Phase Select at any time to
toggle between the Main Output and the Power Analyzer input. The left-most digit of
the front panel display will indicate either "φ1" for the Main Output, or "φ2" for the
Power Analyzer input. If the φ2 front panel display returns a very high frequency
value, it usually means that there is no signal connected to the Power Analyzer input.
Note that the Phase Select key also selects the measurement source to which the
curre nt range setting applies. The current range command is located in the Inp ut menu.
φ2 115V 60Hz

Making Power Analyzer Input Measurements

All measurements are based on acquiring and subsequently processing waveform information. Waveform
information is acquired either from the ac source output or from the Power Analyzer input. When the ac
source is on, it continually takes measurements and updates the front panel meter from whichever
measurement source is active. The Meter key accesses the measurement functions from the front panel.
This menu applies to both the ac source output and the Power Analyzer input.
NOTE: The Shift Harmonic functions also apply to the Power Analyzer input.
If the display indicates OVLD, the measurement capability of the unit has been
exceeded.
Use the Meter menu for making Power Analyzer input measurements:
Action Display
1. Press Phase Select to toggle to the Power Analyzer input. A φ2 will appear
on the display when the Power Analyzer input is selected.
φ2 115V 3.04A
2. Press Meter and q repeatedly to access the following measurement functions:
rms voltage a nd frequency (the default) φ2 <reading>V <reading>Hz
rms voltage and rms current φ2 <reading>V <reading>A
rms curre nt and freque ncy (the default) φ2 <reading>A <reading>Hz
rms voltage and power φ2 <reading>V <reading>W
current crest factor φ2 <reading> CREST F
peak current, repetitive φ2 <reading>A PK REP
peak current, non-repetitive φ2 <reading>A PK NR
apparent power φ2 <reading> VA
reactive power φ2 <reading> VAR
power factor φ2 <reading> PFACTOR
The ac source uses the output frequency value when calculating many of the output measurements. For ac
source output measurements, the ac source uses the programmed output frequency value. For the Power
Analyzer input , the ac source uses the frequency measured at the V SENSE input channel. This is the
default setting. You can also select the I SENSE input channel as the frequency measurement source.
Use the Input menu for selecting the frequency source:
Action Display
If a volta ge signal is not availabl e to measure the output frequency, you must
select the current measurement as the frequency source. Press Input, scroll to
FREQ:SRC VOLT, select the CURR parameter and press Enter.
FREQ:SRC CURR