Power Analyzer Voltage Measurement Accuracy

This test verifies the accuracy of the Power Analyzer voltage measurements. The Phase 2 annunciator on the left side of the display, must be on for the front panel to display Power Analyzer input measurements. Figure B-1A shows the setup. Measure the ac output voltage directly at the output terminals.

Action

1.Make sure the ac source is turned off. Connect the DVM to the output terminals the ac source. Connect the Power Analyzer voltage sense terminals (V SENSE) to the Main Output terminals of the ac source (see figure B-1A).

2.Turn on the ac source with no load. In the Output menu, execute the *RST command to reset the unit to its factory default state.

3.If the Phase 2 annunciator does not appear on the right side of the display, press the Phase Select key.

4.Program the output voltage to 150 volts and set the output current limit to its maximum value.

5.Enable the output by pressing Output On/Off.

6.Record voltage readings at the DVM and on the front panel display.

7.Program the output voltage to 300 volts.

8.Record voltage readings at the DVM and on the front panel display.

Normal Result

*RST

φ2 0 V 60 Hz

CV annunciator on.

Output voltage near 150 V.

φ2 front panel reading within 145 mV of DVM reading .

Output voltage near 300 V.

φ2 front panel reading within

190 mV of DVM reading

Power Analyzer Current Measurement Accuracy

This test verifies the accuracy of the Power Analyzer current measurements. The Phase 2 annunciator on the left side of the display, must be on for the front panel to display Power Analyzer input measurements. Figure B-1A shows the setup. Use the Agilent supplied current shunt. It has the accuracy specified in table B-1. Use wire of sufficient size to carry the maximum rated current of the ac source.

NOTE: Check to make sure that the internal shunt value in the ac source matches the value written on the external current shunt. Refer to "Check the Shunt Value Setting".

Action

Normal Result

1.Turn off the ac source. Connect the 30 ohm load resistor, current shunt, and the DVM across the current shunt. Connect the Power Analyzer current terminals (I SENSE) to the output of the function generator (see figure B-1A).

2.Turn on the ac source. In the Output menu, execute the *RST command to reset the unit to its factory default state.

3.If the Phase 2 annunciator does not appear on the right side of the display, press the Phase Select key.

4.Program the output voltage to 100 volts and set the current limit to 3 amperes. Enable the output by pressing Output On/Off.

5.Record the DVM voltage reading and calculate the rms current. Divide the DVM reading by the current monitor resistor value. Record the front panel reading.

*RST

φ2 0 V 60 Hz

CCannunciator on. Output current near 3 amperes

φ2 front panel reading within

11.5mA of measured output current.

24

Page 24
Image 24
Agilent Technologies 5964-8134 Power Analyzer Voltage Measurement Accuracy, Power Analyzer Current Measurement Accuracy

5964-8134 specifications

Agilent Technologies 5964-8134 is an advanced analytical instrument designed for versatile applications in laboratory settings. It serves as a crucial component for professionals working in fields such as forensic science, environmental testing, and pharmaceutical analysis.

One of the primary features of the 5964-8134 is its exceptional chromatographic performance. The device employs sophisticated separation techniques, utilizing a high-efficiency column that enhances the resolution of complex mixtures. This capability allows scientists to identify and quantify trace levels of compounds in samples with high precision.

Equipped with a state-of-the-art detector, the 5964-8134 offers a broad dynamic range to accommodate various analyte concentrations. This feature ensures that both minor and major components can be accurately measured, making it ideal for applications involving diverse sample matrices. The integrated software provides users with easy access to data analysis tools, enabling seamless interpretation of results.

The 5964-8134 is built on cutting-edge technology, including advanced mass spectrometry which enhances its sensitivity and specificity. This technology enables the detection of low-abundance compounds in complex samples, providing insights that would be difficult to obtain using traditional methods. Additionally, the instrument supports multiple ionization techniques, allowing users to tailor the analysis to specific requirements of their project.

Another characteristic of the Agilent 5964-8134 is its user-friendly interface. The equipment is designed with the operator in mind, featuring intuitive controls and a touchscreen display that promotes ease of use. This design minimizes the learning curve for new users, enabling efficient operation and streamlined workflow in busy laboratory environments.

Moreover, the 5964-8134 ensures robust performance with its advanced temperature control and reliable fluidic systems. These features contribute to improved reproducibility and stability during analyses, which are critical in achieving consistent results across experiments.

With its combination of high performance, cutting-edge technologies, and user-centric design, Agilent Technologies 5964-8134 stands out as a leading choice for professionals seeking reliable and accurate analytical solutions. As laboratories continue to demand innovative instruments that push the boundaries of research capabilities, the 5964-8134 proves to be an essential tool in scientific discovery and analysis.