Calibration Procedure

Table B-1 lists the equipment required for calibration. Figure B-1 shows the test setup. The following procedures assume you understand how to operate front panel keys (see chapter 4).

WARNING LETHAL VOLTAGES. Ac sources can supply 425 V peak at their output. DEATH on contact may result if the output terminals or circuits connected to the output are touched when power is applied. These procedures must be performed by a qualified electronics technician or engineer trained on this equipment.

Summary of Front Panel Calibration Menu

The Entry keypad is used for calibration functions.

Shift

Cal

Press this key to access the calibration menu.

Display

Command Function

CAL ON <value>

Turns calibration mode on when the correct password

 

value is entered.

CAL OFF

Turns calibration mode off

CAL:LEV <char>

Advance to next step in sequence (P1, P2, P3, or P4).

CAL:DATA <value>

Input a calibration measurement.

SHUNT <value>

Enter a value for the external current shunt in ohms.

CAL:VOLT:OFFSET

Begin voltage offset calibration

CAL:VOLT:DC

Begin dc voltage calibration sequence

CAL:VOLT:AC

Begin ac voltage calibration sequence

CAL:VOLT:PROT

Begin voltage protection calibration

CAL:VOLT:EXT

Begin external voltage calibration sequence

CAL:CURR:AC

Begin ac current calibration sequence

CAL:CURR:MEAS

Begin current measurement calibration sequence

CAL:CURR:EXT

Begin external current calibration sequence

CAL:IMP

Begin output impedance calibration sequence

CAL:SAVE

Saves the calibration constants in non-volatile memory.

CAL:PASS <value>

Set new calibration password.

Enable Calibration Mode

Action

1.Reset the unit by selecting Shift, Output, and pressing Enter.

2.Press Shift Calibration, enter the calibration password from Entry keypad, and press Enter. If the password is correct, the Cal annunciator will come on. (If no password has been set, just press Enter.)

If the password is incorrect, an error occurs. If the active password is lost, the calibration function can be recovered by setting an internal switch that defeats password protection (see the Service Manual.)

If CAL DENIED appears, then an internal switch has been set to prevent the calibration from being changed (see the Service Manual.)

Display

*RST

CAL ON 0.0

OUT OF RANGE

CAL DENIED

25

Page 25
Image 25
Agilent Technologies 5964-8134 Calibration Procedure, Summary of Front Panel Calibration Menu, Enable Calibration Mode

5964-8134 specifications

Agilent Technologies 5964-8134 is an advanced analytical instrument designed for versatile applications in laboratory settings. It serves as a crucial component for professionals working in fields such as forensic science, environmental testing, and pharmaceutical analysis.

One of the primary features of the 5964-8134 is its exceptional chromatographic performance. The device employs sophisticated separation techniques, utilizing a high-efficiency column that enhances the resolution of complex mixtures. This capability allows scientists to identify and quantify trace levels of compounds in samples with high precision.

Equipped with a state-of-the-art detector, the 5964-8134 offers a broad dynamic range to accommodate various analyte concentrations. This feature ensures that both minor and major components can be accurately measured, making it ideal for applications involving diverse sample matrices. The integrated software provides users with easy access to data analysis tools, enabling seamless interpretation of results.

The 5964-8134 is built on cutting-edge technology, including advanced mass spectrometry which enhances its sensitivity and specificity. This technology enables the detection of low-abundance compounds in complex samples, providing insights that would be difficult to obtain using traditional methods. Additionally, the instrument supports multiple ionization techniques, allowing users to tailor the analysis to specific requirements of their project.

Another characteristic of the Agilent 5964-8134 is its user-friendly interface. The equipment is designed with the operator in mind, featuring intuitive controls and a touchscreen display that promotes ease of use. This design minimizes the learning curve for new users, enabling efficient operation and streamlined workflow in busy laboratory environments.

Moreover, the 5964-8134 ensures robust performance with its advanced temperature control and reliable fluidic systems. These features contribute to improved reproducibility and stability during analyses, which are critical in achieving consistent results across experiments.

With its combination of high performance, cutting-edge technologies, and user-centric design, Agilent Technologies 5964-8134 stands out as a leading choice for professionals seeking reliable and accurate analytical solutions. As laboratories continue to demand innovative instruments that push the boundaries of research capabilities, the 5964-8134 proves to be an essential tool in scientific discovery and analysis.