Program Listing for "A" Modules

10Load=705

20Chan= 1

30OUTPUT Load;"CHAN";Chan;";CAL ON"

40Cal_curr(Load,Chan,Hi_curr_rng,Hi_curr_hipt,Hi_curr_lopt,l);

50Cal_curr(Load,Chan,Lo_curr_rngLo_cu,rr_hipt,Lo_curr_lopt,0)

60Cal_volt(Load,Chan,Volt_hipt,Volt_lopt)

70Cal_res(Load,Chan,Lo_res_rng,Lo_res_hipt,Lo_res_lopt,0)

80Cal_res.(Load,Chan,Mid_res_rng,Mid_res_hipt,Mid_res_lopt,l)

90Cal_res(Load,Chan,Hi_res_rng,Hi_res_hipt,Hi_res lopt,l),

100OUTPUT Load;l"CAL:SAVE"

110OUTPUT Load;"CAL OFF"

120END

130!

140SUB Cal_curr(Load,Chan,Curr_rng,Curr_hipt,Curr_lopt,Flag)

150PRINT "CURRENT CALIBRATION, RANGE ";Curr_rng

160PRINT "Set power supply according to module calibration table"

170PRINT "Use the correct current shunt for the range you are calibrating"

180PRINT "Press CONTINUE when ready"

190PAUSE

200OUTPUT Load;"CHAN";Chan

210OUTPUT Load;"MODE:CURR"

220OUTPUT Load;"CURR:RANG";Curr_rng

230OUTPUT Load;"CURR";Curr_hipt

240IF Flag THEN WAIT 30

250INPUT "Enter current through shunt for high point in amps",Hipt_curr

260OUTPUT Load;"CAL:LEV:HIGH";Hipt_curr

270IF Flag THEN OUTPUT Load;"CAL:MEAS:HIGH";Hipt_curr

280OUTPUT Load;"CURR";Curr_lopt

290INPUT "Enter current through shunt for low point in amps",Lopt_curr

300OUTPUT Load;"CAL:LEV:LOW";Lopt_curr

310IF Flag THEN OUTPUT Load;"CAL:MEAS:LOW";Lopt_curr

320PRINT "Test unit to verify that main (and readback) values are in spec"

330PRINT "Press CONTINUE when ready to calibrate transient level"

340PAUSE

350OUTPUT Load;"CURR";Curr_lopt

360OUTPUT Load;"TRAN:MODE TOGG"

370OUTPUT Load;"TRIG:SOUR BUS"

380OUTPUT Load;"TRAN ON"

390OUTPUT Load;"CURR:TLEV";Curr_hipt

400OUTPUT Load;"*TRG"

410IF Flag THEN WAIT 30

420INPUT "Enter current through shunt for transient point in amps",Trans_curr

430OUTPUT Load;"CAL:TLEV";Trans_curr

440OUTPUT Load;"TRAN OFF"

450PRINT "Test unit to verify that transient values are in spec"

460PRINT "Press CONTINUE when ready to calibrate next range or mode"

470PAUSE

480SUBEND

490!

500SUB Cal_volt(Load,Chan,Volt_hipt,Volt_lopt)

Calibration 81

Page 81
Image 81
Agilent Technologies 6051A, 6050A manual Program Listing for a Modules, Pause Subend

6051A, 6050A specifications

Agilent Technologies has long been a leader in providing high-performance test and measurement solutions, and the 6050A and 6051A models exemplify this commitment to quality and innovation. The 6050A and 6051A are versatile signal generators that cater to a diverse range of applications, including research and development, manufacturing, and education, making them essential tools in laboratories and production environments.

The Agilent 6050A is a high-performance RF signal generator known for its frequency range capabilities, which span from 100 kHz to 20 GHz. It offers exceptional phase noise performance and low harmonic distortion, making it ideal for applications that require high signal integrity. The device supports various modulation formats, including AM, FM, and pulse modulation, allowing users to generate a wide range of test signals to simulate real-world conditions.

The 6051A builds upon the robust features of the 6050A with enhanced specifications and additional functionalities. It features a larger frequency modulation bandwidth, pushing the envelope for applications requiring more complex signal generation. The 6051A showcases a superior output power range, ensuring that test signals can be reliably produced at varying power levels. This model also includes advanced output control options that allow for precise signal manipulation, making it particularly suited for testing amplifiers and other RF components.

Both models share core technologies that ensure reliable performance, such as direct digital synthesis (DDS) and phase-locked loop (PLL) architectures. These technologies contribute to the exceptional frequency stability and accuracy that engineers and scientists have come to rely on. Additionally, the user-friendly interface integrated into both models simplifies operation and allows for quick configuration changes, facilitating efficient research and testing workflows.

With comprehensive connectivity options, including GPIB, USB, and Ethernet, the 6050A and 6051A can easily integrate into automated test environments. Their reliability, performance, and flexibility make them a perfect choice for those looking to advance their testing capabilities, whether in academic research, product development, or quality assurance in manufacturing.

In summary, the Agilent Technologies 6050A and 6051A signal generators are powerful tools designed to meet the demands of modern RF testing. Their advanced features, paired with Agilent’s reputation for quality and precision, make them invaluable assets in any engineering or research portfolio. Whether you require sophisticated signal generation for prototype testing or educational purposes, these models will deliver the performance needed to support your objectives.