Table 3-2. Maximum Wire Lengths to Limit Voltage Drops

Wire Size

Cross Section

Resistivity

Maximum Length in Meters (Feet) to Limit

Voltage Drop to 0.5 V or Less

AWG Area in mm2

22

0.5

20

0.75

18

1

16

1.5

14

2.5

12

4

10

6

8

10

6

16

4

Ω/kft Ω/km

16.15

40.1

10.16

26.7

6.388

20.0

4.018

13.7

2.526

8.21

1.589

5.09

0.9994

3.39

0.6285

1.95

0.3953

1.24

0.2486

5 A

(6)

2.5

(9.5)

3.7

(15.5)

5.0

(24.5)

7.3

(39.5)

12.2

(62.5)

19.6

(100)

29

(159)

51

(252)

80

(402)

10A

(3)

1.2

(4.5)

1.9

(7.5)

2.5

(12)

3.6

(19.5)

6.1

(31)

9.8

(50)

14.7

(79)

25

(126)

40

(201)

20A

(1.5)

0.6

(2)

0.9

(3.5)

1.3

(6)

1.8

(9.5)

3.0

(15.5)

4.9

(25)

7.4

(39.5)

12.8

(63)

20

(100)

30A

(1)

0.4

(1.5)

0.6

(2.5)

0.8

(4)

1.2

(6.5)

2.0

(10.5)

3.3

(17)

4.9

(27)

8.5

(40)

13.4

(68)

40A (0.77) 0.31 (1.23) 0.47 (2.0) 0.63 (3.1) 0.91 (4.9) 1.52 (7.9) 2.46 (12.5) 3.69 (19.9) 6.41 (31.6) 10.08 (50.37)

50A (0.62) 0.25 (0.98) 0.37 (1.57) 0.50 (2.49) 0.73 (3.46) 1.22 (6.29) 1.96 (10.00) 2.95 (15.91) 5.13 (25.30) 8.06 (40.23)

60 A

(0.52)

0.21

(0.82)

0.31

(1.30)

0.42

(2.07)

0.61

(3.30)

1.01

(5.24)

1.64

(8.34)

2.96

(13.25)

4.27

(21.07)

6.72

(33.51)

Zero-Volt Loading Connections

As shown in Figure 3-15, the Electronic Load can be connected in series with a voltage source or auxiliary power supply greater than 3 V so that the Electronic Load can test devices at its full current capacity down to a zero-volt level. Remote sensing is recommended for improved load regulation and when turning the short on.

Installation 45