Agilent Technologies 6050A, 6051A manual If Flag then 1270 Output @LdRESReshipt 1280

Models: 6051A 6050A

1 98
Download 98 pages 26.44 Kb
Page 90
Image 90

Program Listing for "B" Modules (continued)

1040

!

1050

SUB Cal_res(@Ld,Chan,Res_rng,Res_hipt,Res_lopt,Flag)

1060

PRINT "RESISTANCE CALIBRATION, RANGE";Res_rng

1070

PRINT "Set power supply to calibration information table"

1080

PRINT "Press CONT when ready to continue"

1090

PAUSE

1100

OUTPUT @Ld;"CHAN";Chan

1110

OUTPUT @Ld;"MODE:RES"

1120

OUTPUT @Ld;"RES:RANG";Res_rng

1130

OUTPUT @Ld;"RES";Res_hipt

1140

INPUT "Enter voltage across input terminals in volts",Hipt_volt

1150

INPUT "Enter current through current shunt in amps”,Hipt_curr

1160

Hipt_res=Hipt_volt/Hipt_curr

1170

OUTPUT @Ld;"CAL:LEV:HIGH";Hipt_res

1180

OUTPUT @Ld;"RES";Res_lopt

1190

INPUT "Enter voltage across input terminals in volts",Lopt_volt

1200

INPUT "Enter current through current shunt in amps",Lopt_curr

1210

Lopt_res=Lopt_volt/Lopt_curr

1220

OUTPUT @Ld;"CAL:LEV:LOW;Lopt_res

1230

PRINT "Test unit to verify resistance values"

1240

PRINT "Press CONT when ready to calibrate transient level"

1250

PAUSE

1260

IF Flag THEN

1270

OUTPUT @Ld;"RES";Res_hipt

1280

ELSE

1290

OUTPUT @Ld;"RES";Res_lopt

1300

END IF

1310

OUTPUT @Ld;"TRAN:STAT ON;MODE TOGG;:TRIG:SOUR BUS"

1320

IF Flag THEN

1330

OUTPUT @Ld;"RES:TLEV";Res_lopt

1340

ELSE

1350

OUTPUT @Ld;"RES:TLEV";Res_hipt

1360

END IF

1370

OUTPUT @Ld;"*TRG"

1380

INPUT "Enter voltage across input terminals in volts",Tran_volt

1390

INPUT "Enter current through current shunt in amps",Tran_curr

1400

Tran_res=Tran_volt/Tran_curr

1410

OUTPUT @Ld;"CAL:TLEV";Tran_res

1420

OUTPUT @Ld;"TRAN OFF"

1430

PRINT "Test unit to verify transient values are in spec"

1440

PRINT "Press CONT when ready to end program or calibrate next range"

1450

PAUSE

1460

SUBEND

90 Calibration

Page 90
Image 90
Agilent Technologies 6050A, 6051A manual If Flag then 1270 Output @LdRESReshipt 1280

6051A, 6050A specifications

Agilent Technologies has long been a leader in providing high-performance test and measurement solutions, and the 6050A and 6051A models exemplify this commitment to quality and innovation. The 6050A and 6051A are versatile signal generators that cater to a diverse range of applications, including research and development, manufacturing, and education, making them essential tools in laboratories and production environments.

The Agilent 6050A is a high-performance RF signal generator known for its frequency range capabilities, which span from 100 kHz to 20 GHz. It offers exceptional phase noise performance and low harmonic distortion, making it ideal for applications that require high signal integrity. The device supports various modulation formats, including AM, FM, and pulse modulation, allowing users to generate a wide range of test signals to simulate real-world conditions.

The 6051A builds upon the robust features of the 6050A with enhanced specifications and additional functionalities. It features a larger frequency modulation bandwidth, pushing the envelope for applications requiring more complex signal generation. The 6051A showcases a superior output power range, ensuring that test signals can be reliably produced at varying power levels. This model also includes advanced output control options that allow for precise signal manipulation, making it particularly suited for testing amplifiers and other RF components.

Both models share core technologies that ensure reliable performance, such as direct digital synthesis (DDS) and phase-locked loop (PLL) architectures. These technologies contribute to the exceptional frequency stability and accuracy that engineers and scientists have come to rely on. Additionally, the user-friendly interface integrated into both models simplifies operation and allows for quick configuration changes, facilitating efficient research and testing workflows.

With comprehensive connectivity options, including GPIB, USB, and Ethernet, the 6050A and 6051A can easily integrate into automated test environments. Their reliability, performance, and flexibility make them a perfect choice for those looking to advance their testing capabilities, whether in academic research, product development, or quality assurance in manufacturing.

In summary, the Agilent Technologies 6050A and 6051A signal generators are powerful tools designed to meet the demands of modern RF testing. Their advanced features, paired with Agilent’s reputation for quality and precision, make them invaluable assets in any engineering or research portfolio. Whether you require sophisticated signal generation for prototype testing or educational purposes, these models will deliver the performance needed to support your objectives.