Chapter 2 Initial Operation
Output Checkout
30
Current Output CheckoutThe following steps check basic current functions with a short across the
power supply’s output.
1Turn on the power supply.
The power supply will go into the power-on / reset state; the output is disabled
(the OFF annunciator turns on); the 15V/7A range is selected (the 15V
annunciator turns on); and the knob is selected for voltage control.
2Connect a short acr oss (+) and (-) output t erminals with an insul ated
test lead.
3Enable the output.
The OFF annunciator turns off and the 15V, OVP, and OCP annunciators are lit.
The CV or CC annunciator turns on depending on the resistance of the test
lead. The blinking digit can be adjusted by turning the knob. Notice that the
display is in the meter mode. ‘‘Meter mode’’ means that the display shows the
actual output voltage and current.
4Adjust the voltage limit value to 1.0 volt.
Set the display to the limit mode (the Limit annunciator will be blinking).
Adju st th e vo lta ge l imi t to 1.0 vol t to ass ure CC o per ati on. T he CC annunciator
will turn on.
5Check that the front-panel ammeter properly responds to knob control
for the 15V/7A range.
Set the knob to the current control, and turn the knob clockwise or counter
clockwise when the display is in the meter mode (the Limit annunciator is off).
Check that the ammeter responds to knob control and the voltmeter indicates
nearly zero (the voltmeter will show the voltage drop caused by the test lead).