Chapter 3 Front-Panel Operation
Programming Overcurrent Protection
46
5Exit the OCP menu.
The “ CHA NGE D” me ssa ge i s di spl aye d fo r a s eco nd t o sh ow th at t he n ew O CP
trip level is now in effect. If the OCP settings are not changed, “NO CHANGE”
will be displayed. The power supply will exit the OCP menu and the display
will return to the meter mode. Check that the OCP annunciator turns on.
Checking OCP Operation
To check OCP operation, raise the output current to near the trip point. Then
very gradually increase the output by turning the knob until the OCP circuit
trips. This will cause the power supply’s output current to drop to zero and the
OCP annunciator to blink. The “OCP TRIPPED” message also appears on the
display.
Clearing the Overcurrent Condition
When the OCP condition occurs (the “OCP TRIPPED” message is shown on
the display), the OCP annunciator blinks. When it was caused by external
voltage source such as a battery, disconnect it first. The following steps show
how to clear the overcurrent conditions and get back to normal mode
operation. In the follo wing steps, the displa y will go back to "OCP TRIP PED"
if you let the display time out after about several seconds.
or
1Readjust the OCP trip level or the output current level.
Lower the output current level below the OCP trip point after pressing the
or raise the OCP trip level by using the knob after pressing
the key.
2Move to the clear mode.
You will see the above message after pressing the key. If you
changed the output current level, press the key twice. Turn the
knob to the right until the "OCP CLEAR" appears on the display.
CHANGED
ocp on
Over Current
Over Current
Display Limit
Display Limit
Over Current
Over Current
Over Current