ESG Family Signal Generators Assembly Replacement
AT1 Mechanical Attenuator
Service Guide 5-49
Replacement
1. Reverse the removal procedure and do the following:
When reinstalling the electronic attenuator/RPP assembly, remember to insert its
hinged tabs into the slots located at the base of the chassis wall on which the
assembly will be mounted.
Torque all semi-rigid cables to 9 in-lbs.
Torque all screws to 9 in-lbs.
Figure 5-29 AT1 Mechanical Attenuator (1 of 3)