Agilent Technologies E4438C manual Enhanced multitone1, Awgn

Models: E4438C

1 40
Download 40 pages 3.14 Kb
Page 23
Image 23
Enhanced multitone1

Specifications for Signal Personality Characteristics

Enhanced multitone1

Number of tones

2 to 1024

[arbitrary waveform mode]

 

 

 

Tone spacing

1 kHz to 50 MHz, limited by 80 MHz I/Q bandwidth

[Option 408]

 

 

 

 

Tone power (relative)

0 to –50 dB

 

 

 

 

 

Phase distribution

Fixed, random or parabolic

 

 

 

 

 

Suppression level

–50 to –90 dBc, depending on number of tones

 

 

and available calibration time.

 

 

Expected suppression = 80 dBc –10 log [N/8],

 

 

where N is the number of tones

 

 

 

 

 

Calibration interval

8 hours

 

 

 

 

 

Calibration time

10 minutes (8 tones, –80 dBc suppression)

 

 

 

 

 

Temperature stability

1 dB/°C (typical for IMD products) 5 dB/°C

 

 

(worst case for LO feedthrough and unbalanced

 

 

images)

 

 

 

 

 

 

 

 

 

 

 

 

Enhanced multitone signal with correction applied

AWGN

[real-time mode] [Option 403]

Crest factor [output power set at least 16 dB below maximum power] > 16 dB

Randomness

89 bit pseudo-random generation, repetition period

 

3 x 109 years

Carrier to noise ratio

Magnitude error 0.2 dB at baseband I/Q outputs.

 

 

1. All values typical.