Specifications for Frequency and Power Characteristics

Output power

1.Quoted specifications for 23 °C ± 5 °C. Accuracy degrades by less than 0.03 dB/°C over full temperature range. Accuracy degrades by 0.3 dB above +7 dBm, and by 0.8 dB above +10 dBm.

2.Parentheses denote typical performance.

3.Quoted specifications for 23 °C ± 5 °C. Accuracy degrades by less than 0.03 dB/°C over full temperature range. Accuracy degrades by 0.2 dB above +10 dBm, and by 0.8 dB above +13 dBm.

4.Quoted specifications for 23 °C ± 5 °C. Accuracy degrades by less than 0.02 dB/°C over full temperature range. Accuracy degrades by 0.2 dB above +7 dBm.

Power

 

Option 501-504

With Option UNB

Option 506

250 kHz to 250 MHz

+11 to –136 dBm

+15 to –136 dBm

+12 to –136 dBm

> 250 MHz to 1 GHz

+13 to –136 dBm

+17 to –136 dBm

+14 to –136 dBm

> 1 to 3 GHz

+10 to –136 dBm

+16 to –136 dBm

+13 to –136 dBm

> 3 to 4 GHz

+7 to –136 dBm

+13 to –136 dBm

+10 to –136 dBm

> 4 to 6 GHz

N/A

N/A

+10 to –136 dBm

Typical maximum available power

 

26

 

 

 

 

 

 

 

24

 

Option UNB

 

 

 

 

 

22

 

 

 

Option 506

 

 

 

 

 

 

 

 

 

 

 

 

 

 

[dB]

20

 

 

 

 

 

 

 

 

 

 

 

 

 

Power

18

 

 

 

 

 

 

16

 

 

 

 

 

 

 

 

 

 

 

 

 

 

14

Option 501-504

 

 

 

 

 

 

 

 

 

 

 

12

 

 

 

 

 

 

 

10

 

2000

 

 

 

 

 

0

1000

3000

4000

5000

6000

Frequency [MHz]

Level resolution

 

0.02 dB

 

 

 

 

 

 

 

Level range with Attenuator Hold active

 

 

 

 

 

Option 501-504 With Option UNB

Option 506

 

250 kHz to 1 GHz

23 dB

27 dB

 

24 dB

 

> 1 to 3 GHz

20 dB

26 dB

 

23 dB

 

> 3 to 4 GHz

17 dB

23 dB

 

20 dB

 

> 4 to 6 GHz

N/A

N/A

 

20 dB

 

 

 

 

 

 

Level accuracy [dB]

 

 

 

 

 

Option 501-504 1,2

 

 

 

 

 

 

 

 

Power level

 

 

 

 

+7 to

–50 to

–110 to

< –127 dBm

 

 

 

–50 dBm

–110 dBm

–127 dBm

 

250 kHz to 2.0 GHz

±0.5

±0.5

±0.7

(±1.5)

 

2.0 to 3 GHz

±0.6

±0.6

±0.8

(±2.5)

 

3 to 4 GHz

±0.7

±0.7

±0.9

(±2.5)

 

 

 

 

 

 

 

With Option UNB 2 , 3

 

 

 

 

 

 

 

 

Power level

 

 

 

 

+10 to

–50 to

–110 to

< –127 dBm

 

 

 

–50 dBm

–110 dBm

–127 dBm

 

250 kHz to 2.0 GHz

±0.5

±0.7

±0.8

(±1.5)

 

2.0 to 3 GHz

±0.6

±0.8

±1.0

(±2.5)

 

3 to 4 GHz

±0.8

±0.9

±1.3

(±2.5)

 

 

 

 

 

 

 

Option 506 2, 4

 

 

 

 

 

 

 

 

Power level

 

 

 

 

+7 to

–50 to

–110 to

< –127 dBm

 

 

 

–50 dBm

–110 dBm

–127 dBm

 

250 kHz to 2.0 GHz

±0.6

±0.8

±0.8

(±1.5)

 

2.0 to 3 GHz

±0.6

±0.8

±1.0

(±2.5)

 

3 to 4 GHz

±0.8

±0.9

±1.5

(±2.5)

 

4 to 6 GHz

±0.8

±0.9

(±1.5)

 

6

Page 6
Image 6