Operation Verification

Checking the Output Power

Table 3-4

Leveled Output Power Limits, E8257D

E8257D

Option 520

Frequency

Amplitude

Limits

 

 

 

 

 

Option

dBm

dB

 

 

 

 

250 kHz

None

13

±0.6

 

 

 

 

 

1EA

16

±0.6

 

 

 

 

 

1E1

11

±0.6

 

 

 

 

 

1E1 & 1EA

15

±0.6

 

 

 

 

3.2 GHz

None

13

±0.8

 

 

 

 

 

1EA

16

±0.8

 

 

 

 

 

1E1

11

±0.8

 

 

 

 

 

1E1 & 1EA

15

±0.8

 

 

 

 

20 GHz

None

13

±0.8

 

 

 

 

 

1EA

20

±0.8

 

 

 

 

 

1E1

11

±0.8

 

 

 

 

 

1E1 & 1EA

18

±0.8

 

 

 

 

Option 540

Frequency

Amplitude

Limits

 

 

 

 

 

Option

dBm

dB

 

 

 

 

250 kHz

None

9

±0.6

 

 

 

 

 

1EA

15

±0.6

 

 

 

 

 

1E1

7

±0.6

 

 

 

 

 

1E1 & 1EA

14

±0.8

 

 

 

 

3.2 GHz

None

9

±0.8

 

 

 

 

 

1EA

15

±0.8

 

 

 

 

 

1E1

7

±0.8

 

 

 

 

 

1E1 & 1EA

14

±0.8

 

 

 

 

20 GHz

None

9

±0.9

 

 

 

 

 

1EA

18

±1.0

 

 

 

 

 

1E1

7

±0.9

 

 

 

 

 

1E1 & 1EA

16

±1.0

 

 

 

 

40 GHz

None

9

±0.9

 

 

 

 

 

1EA

14

±1.0

 

 

 

 

 

1E1

7

±0.9

 

 

 

 

 

1E1 & 1EA

12

±1.0

 

 

 

 