Operation Verification

Checking the Maximum Leveled Power

3. Ensure that the RF OUTPUT connector is connected to a 50load.

4.Ensure that the power level entered corresponds to the value listed in Table 3-1.

5.Refer to Table 3-2for the recommended equipment and measure the output power.

If the measured power level is more than the power level listed in Table 3-1, turn the front panel knob counterclockwise until the measured power level equals the power level in Table 3-1.

6.Note the signal generator front-panel amplitude value and use it to repeat the maximum leveled power check.

If all of the above steps are followed and the check continues to fail, contact the nearest Agilent Technologies call center for assistance. Refer to “Contacting Agilent Sales and Service Offices” on page 16 for contact information.

Table 3-1

Frequency and Power Level Limits

 

 

 

 

 

 

 

 

 

 

 

Instrument Option

 

 

Amplitude

Start

Stop

# Points

 

 

Standard

Option 1EA

Frequency

Frequency

 

 

 

 

 

 

 

 

 

E8257D

 

13 dBm

 

16 dBm

250 kHz

3.2 GHz

50

Option 520

 

 

 

 

 

 

 

 

13 dBm

 

20 dBm

3.21 GHz

20 GHz

200

 

 

 

 

 

 

 

 

 

 

 

E8257D

 

9 dBm

 

15 dBm

250 kHz

3.2 GHz

50

Option 540

 

 

 

 

 

 

 

 

9 dBm

 

18 dBm

3.21 GHz

20 GHz

200

 

 

 

 

 

 

 

 

 

 

 

 

 

9 dBm

 

14 dBm

20.1 GHz

40 GHz

200

 

 

 

 

 

 

 

 

E8257D

 

5 dBm

 

14 dBm

250 kHz

3.2 GHz

50

Option 550 and

 

 

 

 

 

 

 

 

5 dBm

 

16 dBm

3.21 GHz

20 GHz

200

Option 567

 

 

 

 

 

 

 

 

 

 

 

5 dBm

 

14 dBm

20.1 kHz

60 GHz

200

 

 

 

 

 

 

 

 

 

 

5 dBm

 

12 dBm

60.1 GHz

67 GHz

100

 

 

 

 

 

 

 

 

E8267D

 

13 dBm

 

N/A

250 kHz

3.2 GHz

50

Option 520

 

 

 

 

 

 

 

 

18 dBm

 

N/A

3.21 GHz

20 GHz

200

 

 

 

 

 

 

 

 

 

 

 

E8267D

 

12 dBm

 

N/A

250 kHz

3.2 GHz

50

Option 532 and

 

 

 

 

 

 

 

 

14 dBm

 

N/A

3.21 kHz

20 GHz

200

Option 544

 

 

 

 

 

 

 

 

 

 

 

12 dBm

 

N/A

20.1 GHz

40 GHz

200

 

 

 

 

 

 

 

 

 

 

10 dBm

 

N/A

40.1 GHz

44 GHz

50

 

 

 

 

 

 

 

 

Chapter 3

23

Page 27
Image 27
Agilent Technologies E8257D/67D manual Amplitude

E8257D/67D specifications

Agilent Technologies has established itself as a leader in the realm of electronic testing and measurement, and the E8257D/67D signal generators are prime examples of its commitment to innovation and quality. These advanced signal generators are engineered to meet the rigorous demands of both research and development as well as manufacturing applications.

The E8257D/67D offers a frequency range of up to 67 GHz, making it suitable for a wide array of modern applications, including radar, satellite communications, and advanced wireless technologies. This wide frequency coverage is complemented by outstanding spectral purity and phase noise performance, which are crucial for applications that require high signal fidelity.

One of the standout features of the E8257D/67D is its ability to generate a variety of waveforms. In addition to traditional continuous wave signals, it supports frequency sweeping and pulse modulation, allowing users to design complex test scenarios that simulate real-world conditions. The instrument's high-resolution frequency and amplitude settings provide precision control, essential for characterizing high-performance components.

The E8257D/67D employs advanced digital signal processing technology that enhances both its accuracy and stability. This digital architecture allows for superior linearity and minimized distortion, which are vital parameters in many testing scenarios. Furthermore, the generator’s fast-switching capabilities enable rapid changes in frequency and amplitude, facilitating dynamic testing environments.

In terms of connectivity, the E8257D/67D supports various interfaces, including LAN, USB, and GPIB, allowing for seamless integration into automated test systems. This flexibility not only maximizes productivity but also simplifies system integration and data acquisition.

The user-friendly GUI (Graphical User Interface) of the E8257D/67D enhances the user experience, making it easier to navigate and configure settings. With its comprehensive set of built-in functions and measurement options, users can efficiently perform complex tasks without extensive training.

Overall, the Agilent Technologies E8257D/67D signal generators excel in providing high-performance signal generation capabilities. With its impressive specifications, versatility, and ease of use, it is an ideal choice for engineers and technicians seeking to ensure the highest standards in electronic design and testing. Whether in R&D labs or production lines, the E8257D/67D stands out as a reliable tool for advanced signal generation.