Basic Operation

Configuring the RF Output

6.Press the up arrow key.

Each press of the up arrow key increases the frequency by the increment value last set with the Incr Set hardkey. The increment value is displayed in the active entry area.

7.The down arrow decreases the frequency by the increment value set in the previous step. Practice stepping the frequency up and down in 1 MHz increments.

You can also adjust the RF output frequency using the knob. As long as frequency is the active function (the frequency is displayed in the active entry area), the knob will increase and decrease the RF output frequency.

8.Use the knob to adjust the frequency back to 700 MHz.

Setting the Frequency Reference and Frequency Offset

The following procedure sets the RF output frequency as a reference frequency to which all other frequency parameters are relative. The frequency initially shown on the display is 0.00 Hz (the frequency output by the hardware minus the reference frequency). Although the display changes, the frequency output does not change. Any subsequent frequency changes are shown as incremental or decremental to 0 Hz.

1.Preset the signal generator: Press: Preset

2.Set the frequency reference to 700 MHz:

Press: Frequency > 700 > MHz > More (1 of 3) > Freq Ref Set.

This activates the frequency reference mode and sets the current output frequency (700 MHz) as the reference value. The FREQUENCY area displays 0.000 Hz, which is the frequency output by the

hardware (700 MHz) minus the reference value (700 MHz). The REF indicator activates and the Freq Ref Off On softkey toggles to On.

3.Turn on the RF output: Press RF On/Off.

The display annunciator changes from RF OFF to RF ON. The RF frequency at the RF OUTPUT connector is 700 MHz.

4.Set the frequency increment value to 1 MHz: Press Frequency > Incr Set > 1 > MHz.

5.Increment the output frequency by 1 MHz: Press the up arrow key.

The FREQUENCY area display changes to show 1.000 000 000 MHz, which is the frequency output by the hardware (700 MHz + 1 MHz) minus the reference frequency (700 MHz). The frequency at the RF OUTPUT changes to 701 MHz.

6.Enter a 1 MHz offset: Press More (1 of 3) > Freq Offset > 1 > MHz.

The FREQUENCY area displays 2.000 000 00 MHz, which is the frequency output by the hardware (701 MHz) minus the reference frequency (700 MHz) plus the offset (1 MHz). The OFFS indicator activates. The frequency at the RF OUTPUT connector is still 701 MHz.

Chapter 2

29

Page 39
Image 39
Agilent Technologies E8267C PSG, E8247C PSG CW, E8257C PSG manual Setting the Frequency Reference and Frequency Offset

E8267C PSG, E8247C PSG CW, E8257C PSG specifications

Agilent Technologies, a leading provider of electronic test and measurement equipment, offers a range of signal generators that are pivotal in various applications, including telecommunications, aerospace, and research. Among their top-tier offerings are the E8257C PSG, E8247C PSG CW, and E8267C PSG. These signal generators are known for their exceptional performance, flexibility, and cutting-edge technologies.

The E8257C PSG is a high-performance microwave signal generator capable of generating signals from 10 MHz to 40 GHz. Its versatility makes it ideal for applications such as radar and satellite communications. One of the standout features of the E8257C is its low phase noise and superior output power, which ensure high-quality signal generation. The instrument incorporates advanced digital signal processing technology, allowing users to create complex modulated signals with precision. It also boasts a comprehensive set of modulation formats, including AM, FM, PM, and pulse modulation, making it suitable for diverse testing scenarios.

The E8247C PSG CW is designed specifically for continuous wave (CW) applications, covering frequencies from 500 kHz to 20 GHz. This generator is optimized for high stability and low output noise, making it ideal for applications that require a reliable signal source, such as manufacturing testing and maintenance of communication systems. The E8247C features a compact design and is equipped with an intuitive user interface, allowing quick access to essential functions. Its frequency agility and fast switching capabilities enhance productivity, enabling engineers to optimize their workflow.

The E8267C PSG delivers unparalleled performance in the microwave frequency range of 250 kHz to 70 GHz. It includes advanced modulation capabilities, allowing users to generate complex waveforms with relative ease. The E8267C supports various digital modulation schemes, including QAM, which is crucial for testing modern communication systems. Its high output power and excellent spectral purity make it suitable for both R&D and manufacturing applications. Additionally, the E8267C features built-in functionality for frequency and power sweeps, providing users with extensive testing options.

In summary, Agilent Technologies’ E8257C, E8247C, and E8267C signal generators are integral tools for professionals seeking high performance and versatility in their testing environments. With unique features and advanced technologies, these generators cater to the demands of modern applications, ensuring reliable and accurate signal generation for a wide range of industries. Whether it's for research, development, or manufacturing, these instruments demonstrate Agilent's commitment to delivering top-notch testing solutions.