5 - Viewing Test Data
58
Viewing Reports
Reports are available in either Short form or Long form, and can be previewed in the Report window. User
remarks can be added to a test file by selecting the Remarks menu tab. Menu tabs on the bottom of the screen
let you quickly access each area. The Printer button sends your report to the default printer.
NOTE: You must run Post processing to generate data for the long form reports.
If your test report indicates that your test was non-compliant, it may be due to one of the following
conditions:

Causes for Non-compliant EN 61000-3-2 or EN 60555-2 Tests

ñThe results of a pre-test indicate that the voltage distortion of the ac source is "OUT OF SPEC".
ñThe user set the test limit override to a value > 100%.
ñThe user specified a Class A device, but a Class D waveform is detected, the active power is
within Class D boundaries, and the "Motor Driven Device" checkbox is NOT checked.
ñA data collection error was detected.
ñA current limit was detected.
ñA "Protection" error such as OVP was detected.

Causes for Non-compliant EN 61000-3-3 Tests

ñThe user set the integration time (or Observation Period) to a value other than 10 minutes
ñThe user set any or all Pst/Plt or RMS thresholds to a value > 100%
ñA data collection error was detected.
ñA current limit was detected.
ñA "Protection" error such as OVP was detected.