C - Command Summary
92
SENSe:CURRent:PREFerenceThis command sets the phase reference for current harmonic phase measurements. If VOLTage is
selected, the reference is the fundamental component of the measured output voltage. If CURRent is
selected, the reference is the fundamental component of the measured output current.
Command Syntax SENSe:CURRent:PREFerence <CRD>
Parameters VOLTage | CURRent
*RST Value CURRent
Examples SENS:CURR:PREF CURR
Query Syntax SENSe:CURRent:PREF?
Returned Parameters <CRD>
Related Commands ABORt MEAS:ARR:CURR:PHAS
SENSe:WINDowThis command sets the window function which is used in harmonic measurements. The choice of
parameters is affected by the SYSTem:CONF command. If NORMal is selected, HANNing, KBESsel,
or RECTangular may be selected. IF IEC mode is selected, only HANNing and RECTangular may be
selected. KBESsel is the preferred window and should be used for most measurements in NORMal
mode. HANNing and RECTangular are available for making harmonic current measurements that
comply with the regulatory requirements.
Command Syntax SENSe:WINDow [:TYPE] <window>
Parameters HANNing | KBESsel | RECTangular
*RST Value KBESsel
(RECTangular is the default setting when the HFTS software is run)
Examples SENS:WIND RECT
Query Syntax SENSe:WINDow?
Returned Parameters <CRD>
Related Commands MEASure:ARRay:CURRent:HARMonic? SYSTem:CONF
MEASure:CURRent:HARMonic?
MEASure:ARRay:VOLTage:HARMonic?
MEASure:VOLTage:HARMonic?