5.On
NOTE: If this step is not followed, damage can occur to the calibration temperature probe.
6.On
WARNING
NEVER insert hands or objects into basket until it has completely stopped. Doing so could result in serious injury.
SW012
7.Restore input power to the machine.
8.Wait until machine message displays “CY” followed by a
9.Program one of the unused cycles to Fill to High Level using both of the water inlet valves. Select a time limit for the step that would be sufficient to complete the calibration procedure.
10.When machine has completed the Fill to High Level, press and hold the Up (arrow) key on the keypad to display the machine temperature. Make sure that the
11.Using a
Programming
CHM521R
CHM521R
Figure 2
12.When the temperatures match, calibration is completed.
13.Remove power from the machine input.
14.Reconnect any wires/connectors/fuses that were removed, making sure they are installed exactly as they were removed.
15.Reinstall all machine protective covers.
16.Remove the temperature calibration probe from the bottom of the machine basket.
17.Restore input power to machine.
18.Machine is now ready for use.