Analog Devices C1599c07, AD620 specifications Pressure Measurement

Models: AD620 C1599c07

1 16
Download 16 pages 59.35 Kb
Page 12
Image 12

AD620

 

+5V

 

 

 

 

 

 

 

 

7

20kV

 

 

 

 

3

 

 

 

3kV

3kV

 

 

 

 

8

 

 

REF

 

 

 

 

 

 

 

 

 

 

 

 

 

 

G=100

AD620B

6

IN

DIGITAL

3kV

3kV

499V

5

10kV

ADC

DATA

 

 

1

 

 

2

4

 

 

OUTPUT

 

 

 

 

 

 

 

 

 

AD705

AGND

 

 

 

 

 

20kV

 

 

1.7mA

 

 

1.3mA

0.6mA

 

 

 

 

0.10mA

 

 

 

 

 

MAX

MAX

 

 

 

 

 

 

 

 

Figure 35. A Pressure Monitor Circuit which Operates on a +5 V Single Supply

Pressure Measurement

Although useful in many bridge applications such as weigh scales, the AD620 is especially suitable for higher resistance pressure sensors powered at lower voltages where small size and low power become more significant.

Figure 35 shows a 3 kΩ pressure transducer bridge powered from +5 V. In such a circuit, the bridge consumes only 1.7 mA. Adding the AD620 and a buffered voltage divider allows the signal to be conditioned for only 3.8 mA of total supply current.

Small size and low cost make the AD620 especially attractive for voltage output pressure transducers. Since it delivers low noise and drift, it will also serve applications such as diagnostic non- invasive blood pressure measurement.

Medical ECG

The low current noise of the AD620 allows its use in ECG monitors (Figure 36) where high source resistances of 1 MΩ or higher are not uncommon. The AD620’s low power, low supply voltage requirements, and space-saving 8-lead mini-DIP and SOIC package offerings make it an excellent choice for battery powered data recorders.

Furthermore, the low bias currents and low current noise coupled with the low voltage noise of the AD620 improve the dynamic range for better performance.

The value of capacitor C1 is chosen to maintain stability of the right leg drive loop. Proper safeguards, such as isolation, must be added to this circuit to protect the patient from possible harm.

PATIENT/CIRCUIT

 

 

+3V

 

 

 

 

 

 

 

 

 

PROTECTION/ISOLATION

 

 

 

 

 

 

 

R1

R3

 

 

0.03Hz

 

 

C1

10kV

24.9kV

RG

 

 

 

AD620A

HIGH

G = 143

OUTPUT

 

 

 

 

 

R2

8.25kV

 

PASS

 

1V/mV

 

R4

 

G = 7

FILTER

 

 

 

1MV

24.9kV

 

 

 

 

 

 

 

 

 

 

 

OUTPUT

 

 

 

 

 

 

 

AMPLIFIER

 

AD705J

 

 

 

 

 

 

 

 

 

 

–3V

 

 

 

Figure 36. A Medical ECG Monitor Circuit

–12–

REV. E

Page 12
Image 12
Analog Devices C1599c07, AD620 specifications Pressure Measurement