UG-002

Evaluation Board User Guide

 

 

08129-026

Figure 25. Register Debug Window

11.Bit 3 of Register 0x17 must be set to 1 to enable the PLL locked pin as a test port. By default, the 26 MHz crystal frequency is present at the output.

12.The ac signal has a dc offset of approximately 1.6 V. Be aware that some test equipment has a limit of 0 V dc; therefore, a dc block must be used in such cases.

SETTING THE 19.44 MHz MODE

The following instructions can be used to configure the AD9551 to operate in the 19.44 MHz mode. Refer to the AD9551 data sheet for more details about this mode of operation.

The following case is an example for the AD9551 in

19.44 MHz mode:

1.Replace the crystal on the evaluation board.

2.Replace the loop filter capacitor on the evaluation board.

3.Click Click for Details in the PLL & DCXO box in the main window of the evaluation board software to access the Input PLL + DCXO window (see Figure 26).

4.Select 19.44MHz Mode in the Input PLL Mode: box, and then click OK.

5.Click Click for Details in the PFD & Charge Pump box in the main window of the evaluation board software to access the PFD and Charge Pump window (see Figure 27).

08129-027

Figure 26. Input PLL + DCXO Window

08129-028

Figure 27. PFD and Charge Pump Window

6.Change the charge pump current to the appropriate current level specified in the AD9551 data sheet (approximately 25 μA), and then click OK.

7.If the LOAD button in the main window starts blinking, this indicates that there are settings that have not been loaded to the AD9551 evaluation board. Click LOAD to load these settings to the evaluation board.

ESD CAUTION

Evaluation boards are only intended for device evaluation and not for production purposes. Evaluation boards are supplied “as is” and without warranties of any kind, express, implied, or statutory including, but not limited to, any implied warranty of merchantability or fitness for a particular purpose. No license is granted by implication or otherwise under any patents or other intellectual property by application or use of evaluation boards. Information furnished by Analog Devices is believed to be accurate and reliable. However, no responsibility is assumed by Analog Devices for its use, nor for any infringements of patents or other rights of third parties that may result from its use. Analog Devices reserves the right to change devices or specifications at any time without notice. Trademarks and registered trademarks are the property of their respective owners. Evaluation boards are not authorized to be used in life support devices or systems.

©2009 Analog Devices, Inc. All rights reserved. Trademarks and registered trademarks are the property of their respective owners.

UG08129-0-4/09(0)

Rev. 0 Page 12 of 12

Page 12
Image 12
Analog Devices UG-002, AD9551 manual Setting the 19.44 MHz Mode, ESD Caution