Amplitude Modulation (Option 14)

Amplitude Modulation (Option 14)

All amplitude modulation specifications apply at 50% depth, 1 kHz rate, with RF level set 6 dB below maximum specified leveled output power, unless otherwise noted. Amplitude Modulation is not available <10 MHz with Option 22.

AM Depth (typical): 0-90% linear; 20 dB log

AM Bandwidth (3 dB):

DC to 50 kHz minimum

DC to 100 kHz typical

Flatness (DC to 10 kHz rates): ±0.3 dB

Accuracy: ±5%

Distortion: <5% typical

Incidental Phase Modulation (30% depth, 10 kHz rate):

<0.2 radians typical

External AM Input: Log AM or Linear AM input, rear-panel BNC, 50input impedance. For internal modulation, add LF Generator Option 23.

Sensitivity:

Log AM: Continuously variable from 0 dB per volt to 25 dB per volt.

Linear AM: Continuously variable from 0% per volt to 100% per volt.

Maximum Input: ±1V

LF Generator (Option 23)

Two internal waveform generators are added, one providing

a frequency or phase modulating signal and the other an amplitude modulating signal. This Low Frequency (LF) Generator option can only be ordered in combination with either FM/Φ M or AM options, 12 and 14 respectively.

Waveforms: Sinusoid, square-wave, triangle, positive ramp, negative ramp, Gaussian noise, uniform noise. (Check Option 10 for User-Defined)

Rate:

0.1 Hz to 1 MHz sinusoidal

0.1 Hz to 100 kHz square-wave, triangle, ramps

Resolution: 0.1 Hz

Accuracy: Same as instrument timebase

Output: Two BNC connectors on the rear panel, FM/Φ M OUT and

AM OUT

External Pulse Modulation (Option 13)

Pulse modulation specifications apply at maximum rated power, unless otherwise noted. Pulse modulation is not available <10 MHz with Option 22.

On/Off Ratio: >80 dB

Minimum Leveled Pulse Width:

100ns, 2 GHz1

1µs, <2 GHz1

Minimum Unleveled Pulse Width: <10 ns

Level Accuracy Relative to CW (100 Hz to 1 MHz PRF):

±0.5 dB, 1 µs pulse width ±1.0 dB, <1 µs pulse width

Pulse Delay (typical): 50 ns in External Mode

PRF Range:

DC to 10 MHz, unleveled 100 Hz to 5 MHz, leveled

Frequency

Rise & Fall

 

Pulse Width

Video

Range

Time

Overshoot

Compression

Feedthrough

 

(10% to 90%)

 

 

 

10 to

 

 

 

 

<31.25 MHz

400 ns*

33%*

40 ns*

±70 mV*

(Opt. 4)

 

 

 

 

 

 

 

 

 

31.25 to

 

 

 

 

<125 MHz

90 ns*

22%*

12 ns*

±130 mV*

(Opt. 4)

 

 

 

 

 

 

 

 

 

125 to

 

 

 

 

<500 MHz

33 ns*

11%*

12 ns*

±70 mV*

(Opt. 4)

 

 

 

 

 

 

 

 

 

500 to

 

 

 

 

<2200 MHz

15 ns

10%

12 ns*

±15 mV*

(Opt. 4)

 

 

 

 

10 to

15 ns

 

 

 

<1000 MHz

10 ns*

10%

8 ns*

±15 mV*

(Opt. 5)

 

 

 

 

 

 

 

 

 

1 to

10 ns

 

 

 

<2 GHz

5 ns*

10%

8 ns*

±15 mV*

(Opt. 5)

 

 

 

 

2 to

10 ns

10%2

 

 

65 GHz

5 ns*

8 ns*

±10 mV*

External Input: Rear-panel BNC. For internal modulation, add Pulse Generator Option 24.

Drive Level: TTL compatible input

Input Logic: Positive-true or negative-true, selectable from modulation menu.

Pulse Generator (Option 24)

Modes: Free-run, triggered, gated, delayed, singlet, doublet, triplet, quadruplet.

Parameter

Selectable Clock Rate

 

40 MHz

10 MHz

Pulse Width

25 ns to 419 ms

100 ns to 1.6 s

 

 

 

Pulse Period 3

250 ns to 419 ms

600 ns to 1.6s

Variable Delay

 

 

Singlet

0 to 419 ms

0 to 1.6 s

Doublet

100 ns to 419 ms

300 ns to 1.6 s

Triplet

100 ns to 419 ms

300 ns to 1.6 s

Quadruplet

100 ns to 419 ms

300 ns to 1.6 s

 

 

 

Resolution

25 ns

100 ns

 

 

 

Accuracy: 10 ns (5 ns typical)

Inputs/Outputs: Video pulse and sync out, rear-panel BNC connectors

Pulse Generator option is not available without Pulse Modulation Option 13.

12.2 GHz with Option 4, DDC.

2For 50 and 65 GHz units, overshoot >40 GHz is 20% typical at rated power.

3Period must be longer than the sum of delay and width by 5 clock cycles minimum.

* Typical

9