Accuracy and Flatness
Accuracy specifies the total worst case accuracy. Flatness is included within the accuracy specification.
Step Sweep and CW Modes:
Attenuation |
| F r e q u e n c y | ( G H z ) |
|
Below | ≤40 | |||
Max Power | ||||
|
|
|
|
|
Accuracy:➀ |
|
|
|
|
±1.0 dB | ±1.5 dB | ±1.5 dB | ±1.5 dB | |
±1.0 dB | ±1.5 dB | ±3.5 dB➁ | N/A | |
±1.0 dB | ±1.5 dB➁ | ±3.5 dB➁ | N/A | |
|
|
|
|
|
Flatness:➀ |
|
|
|
|
±0.8 dB | ±1.1 dB | ±1.1 dB | ±1.1 dB | |
±0.8 dB | ±1.1 dB | ±3.1 dB➁ | N/A | |
±0.8 dB | ±2.1 dB➁ | ±3.1 dB➁ | N/A | |
|
|
|
|
|
➀0 to 25 dB or to minimum rated power, whichever is higher
➁Typical
Analog Sweep Mode (typical):
Attenuation |
| F r e q u e n c y | ( G H z ) |
|
Below |
|
|
|
|
Max Power | ||||
Accuracy: |
|
|
|
|
±2.0 dB | ±2.0 dB | ±2.0 dB | ±3.0 dB | |
±3.5 dB | ±3.5 dB | ±4.6 dB | ±5.6 dB | |
±4.0 dB | ±4.0 dB | ±5.2 dB | ±6.2 dB | |
±5.0 dB | ±5.0 dB | ±6.2 dB | ±7.2 dB | |
|
|
|
|
|
Flatness: |
|
|
|
|
±2.0 dB | ±2.0 dB | ±2.0 dB | ±2.5 dB | |
±3.5 dB | ±3.5 dB | ±4.1 dB | ±5.1 dB | |
±4.0 dB | ±4.0 dB | ±4.6 dB | ±5.6 dB | |
±5.0 dB | ±5.0 dB | ±5.2 dB | ±6.2 dB | |
|
|
|
|
|
Typical maximum MG3692A available output power
Typical maximum MG3694A available output power
Other Output Power Specifications
Output Units: Output units selectable as either dBm or mV. Selection of mV assumes 50Ω load. All data entry and display are in the selected units.
Output Power Resolution: 0.01 dB or 0.001 mV
Source Impedance: 50Ω nominal
Source SWR (Internal Leveling): <2.0 typical
Power Level Stability with Temperature: 0.04 dB/deg C typical
Level Offset: Offsets the displayed power level to establish a new reference level.
Output On/Off: Toggles the RF output between an Off and On state. During the Off state, the RF oscillator is turned off. The On or Off state is indicated by two LEDs located below the OUTPUT ON/OFF key on the front panel.
RF On/Off Between Frequency Steps: System menu selection of RF On or RF Off during frequency switching in CW, Step Sweep, and List Sweep modes.
RF On/Off During Retrace: System menu selection of RF On or RF Off during retrace.
Internal Leveling: Power is leveled at the output connector in all modes.
External Leveling:
External Detector: Levels output power at a remote detector location. Accepts a positive or negative 0.5 mV to 500 mV input signal from the remote detector. EXT ALC ADJ adjusts the input signal range to an optimum value. BNC connector, rear panel. External Power Meter: Levels output power at a remote power meter location. Accepts a ±1V full scale input signal from the remote power meter. EXT ALC ADJ adjusts the input signal range to an optimum value. BNC connector, rear panel. External Leveling Bandwidth: 30 kHz typical in Detector mode.
0.7Hz typical in Power Meter mode. User Level Flatness Correction:
Number of points: 2 to 801 points per table
Number of tables: 5 available
Entry modes: GPIB power meter or computed data
CW Power Sweep
Range: Sweeps between any two power levels at a single CW frequency.
Resolution: 0.01 dB/step (Log) or 0.001 mV (Linear)
Accuracy: Same as CW power accuracy.
Log/Linear Sweep: Power sweep selectable as either log or linear. Log sweep is in dB; linear sweep is in mV.
Step Size:
Step Dwell Time: Variable from 1 ms to 99 seconds. If the sweep crosses a step attenuator setting, there will be a sweep dwell of approximately 20 ms to allow setting of the step attenuator.
Sweep Frequency/Step Power
A power level step occurs after each frequency sweep. Power level remains constant for the length of time required to complete each sweep.
Internal Power Monitor (Option 8)
Sensors: Compatible with Anritsu
Range: +16 dBm to
Accuracy: ±1 dBm, (+16 to
Resolution: 0.1 dBm minimum | 7 |