AudioQuest M-27 specifications is the percent dose accumulated at 80 dB

Models: M-27

1 25
Download 25 pages 1.77 Kb
Page 15
Image 15
.61 is the percent dose accumulated at 80 dB.

.

.61 is the percent dose accumulated at 80 dB.

1.23 is the percent at 81 dB, etc.

The percent dose statistical distribution is accumulated without a threshold. Of course, the dose calculations in the summary exclude any dose accumulated below the threshold.

The graph is similar to the % TIME graph.

If the internal TEXT or GRAPH switch is off, that portion will not be printed. 3 dB Only Systems

The previous examples illustrated the M-27 when used with two different exchange rates. It can be used as a 3 dB only system with three different histograms, one with no threshold and two with thresholds. It will also print four statistical distributions, two with a Time constant and two without. If the M-27 is used for community or 3 dB industrial monitoring, the printout will contain much more information than would a data logger with a single printout.

NOTE

If the criterion level is set to 85 dB, then the dose reading can be easily converted to Pascal squared hours. For a 3 dB system with a 85 dB criterion level, a 100% dose is equal to 1 Pascal squared hour. To calculate the exposure in Pascal squared hours, divide the dose by 100.

Dump Data

If the CODE/HL3 is cycled until the display shows CPU, raw data may be dumped to a computer when PRINT is pressed. Each line contains an address and 16 bytes of memory starting at that address (see Figure 16).

See Memory Map section for a description of the data.

General When printing, the M-27 remembers which keys have been pressed

Page 15
Image 15
AudioQuest M-27 specifications is the percent dose accumulated at 80 dB