AudioQuest M-27 specifications Sound Level Meter Keys, Dosimeter Keys, Time Keys, Event Keys

Models: M-27

1 25
Download 25 pages 1.77 Kb
Page 4
Image 4
Sound Level Meter Keys

 

In steps 1 to 3 only data for the HTL (High Threshold Level), LTL (Low Threshold

 

Level), or 3 dB (No Thresh-old) will he displayed or printed. In step 4 data for HTL,

 

LTL, and 3 dB will be printed. Step 5 dumps raw data to the Computer. Step 6 reads or

 

resets the Clock. Step 7 reads or stores the Calibrator level.

Sound Level Meter Keys

SOUND

Sound Level in decibels. Will repeat each second until

LEVEL

another key is pressed. To conserve power, press RUN,

 

PAUSE or any other key to exit sound level when it is

 

not needed.

PEAK

The highest unweighted Peak Level in decibels during

LEVEL

Run Time or an Event. In 30 dB range lowest peak

 

measurable is 78 dB; in 50 dB range the lowest measurable is 98 dB.

MAX

The Maximum Level in decibels after the filter and Slow

LEVEL

or Fast time constant circuit.

Dosimeter Keys

 

L-AVG

The average integrated sound level in decibels for the

LEQ

HTL (High Threshold Level), LTL (Low Threshold Level), or the equivalent level for 3

 

dB. It is the accumulated sound averaged during the Run Time.

DOSE

The accumulated Dose in percent for HTL, LTL, or 3 dB.

TWA

Time Weighted Average in decibels for HTL and LTL.

SEL

It is the accumulated sound averaged over 8 hours. For

 

3 dB it is the Sound Exposure Level of the accumulated

 

sound averaged over 1 second.

Time Keys

 

RUN

The total time the unit was running and accumulating

TIME

data. If the total time is less than 30 minutes, the co-

 

lon will blink if the key is held down and the time shown is in minutes and seconds. If the

 

colon stays on, the time shown is in hours and minutes.

PAUSE

The time the unit was on but not accumulating data.

TIME

The colon blinks as in Run Time

.

 

U.L.

The time exposure was above the Upper Limit. The

TIME

colon blinks as in Run Time.

Event Keys

 

The event feature accumulates Lavg/Leq, Dose, TWA/SEL for 3 dosimeters, Peak Level, Max

Level, Run Time, and Start Time for up to 15 different events. Each time a key is pressed in the Event

mode, the display will show the event number and the data.

EVENT

Enters Events and cycles through the Events with

 

data. The display shows the event number, a "CE" if it is the Current Event or else a "-E"

 

for a previous Event.

 

As an example: 12CE means 12 is the Current Event, 3-E means 3 was a previous Event.

 

Only previous events with data will be displayed.

RUN

Starts accumulating data for the Current Event.

PAUSE

Stops accumulating data and cycles to the next Event.

RESET

 

ON/OFF

Exits the Event mode.

Procedure to Accumulate Event Data

 

To enter the Event mode, press EVENT. To accumulate data, press RUN. To stop, press

PAUSE/RESET. Press RUN and PAUSE/RESET again to accumulate another event. Press ON/OFF to exit

events.

 

The following keys can be used to read data in the current event or in any previous event.

CODE

Cycles through HTL, LTL, and 3 dB. Displays the cur-

HL3

rent time or elapsed time for the current event and

 

the start time for any previous event.

Page 4
Image 4
AudioQuest M-27 Sound Level Meter Keys, Dosimeter Keys, Time Keys, Event Keys, Procedure to Accumulate Event Data