Barco MGD 2621 user manual Connection of video signals

Models: MGD 2621

1 27
Download 27 pages 42.71 Kb
Page 10
Image 10

Connection of video signals

The connected equipment must comply to all relevant safety demands.

Important: The use of low-quality video cables can distort the video signal and influence diagnosis.

1.Check the termination impedance of the display’s video input and the impedance of the imaging board that produces the video signals you want to connect. You can find the termination impedance of the video input on the label on the rear of the display. Both values must be equal. If not, the quality of the images on the display will be inferior.

2.Connect the video output of the computer’s imaging board to the video and sync inputs on the display’s rear panel. Use a proper video cable. The video cable is not supplied with the display, unless the display comes as part of a complete BARCO MeDis system, that also contains an imaging board.

The inputs accept the following signals:

Video with separate horizontal and vertical sync.

-The video cable has 3 wires.

-Connect the video signal to the connector Video {6}.

-Connect the horizontal sync signal to the connector HS/CS {7}.

-Connect the vertical sync signal to the connector VS {8}.

Video with external composite sync.

-The video cable has 2 wires.

-Connect the video signal to the connector Video {6}.

-Connect the composite sync signal to the connector HS/CS {7}.

Video with internal composite sync (sync on video).

-The video cable has 1 wire.

-Connect the video (with sync) signal to the connector Video {6}.

Notes:

The video inputs cannot be connected in loop- through (daisy-chain).

The required video amplitude: 700 mV ± 3 dB.

The required sync. amplitude: 400 mV.

Installation10

Page 10
Image 10
Barco MGD 2621 user manual Connection of video signals