NurseCall Relay Unit Programming | en 21
Bosch Security Systems User Manual F.01U.252.722 | V1 | 2011.12
Manual selection of the NurseCall Main Unit
5.a. Scroll with and to set manually the identification number. Confirm with
or the Green button. The Relay Unit signalizes that it is not connected to the RS-485 bus.
5.b. Connect the NurseCall Relay Unit to the RS-485 bus. The identification number is now
displayed at the bottom right.
4.3 Special settings
After pressing three times quickly, you can enter special codes.
4.3.1 Displaying the firmware version
1. Type the code 194155.
2. The version of the firmware will be displayed for a few seconds.
4.3.2 Resetting all the parameters and buffers
1. Type the code 194156.
2. When the unit resets, it plays a short melody and displays a temporary message.
3. After a few seconds, the unit goes back to the first use display.
4. Select the locating mode and the identification number. See Section 4.2 First use.
4.3.3 Assistance and fire priority
This command sets the assistance call and the fire alarm as priority calls.
This means that alarms of these types will be displayed first.
1. Type the code 123991. The unit plays a melody and a confirmation message is displayed.
NurseCall
RELAY UNIT OFF !!!
Enter Code
Software REV A
V2.02 BN111.242.00A
CAUTION!
Disconnect the RS-485 bus before performing a reset. Reconnect it when you are finished.
NOTICE!
The following procedure resets the NurseCall Relay Unit to its original factory settings. It
erases all the parameters, as well as the alarms and events buffers. This reset is mandatory if
you wish to change the locating mode or the identification number.
Please Wait...
ASSISTANCE & FIRE
PRIORITY