5. Basic Photography
-21-
5.3.3 Cobalt Photography Mode
In this mode, photography using the exciter filter for fluorescein angiography is possible.
1Enter the study information into the computer.
2Select the cobalt photography mode.
Press the COBALT switch.
Or select the photography mode with the control
software.
The COBALT lamp turns on. COBALT is displayed on
the observation monitor.
3Perform the procedure starting from step 3 in 5.3.1 Color Photography Mode (see page14).
5.3.4 Fluorescein Angiography Mode
If taking multiple photographs, a mydriatic must be used.
1Enter the study information into the computer.
2Select the fluorescein angiography mode.
Press the FLUO switch.
Or select the photography mode with the control
software.
The FLUO lamp turns on. FLUO is displayed on the
observation monitor.
3Disinfect the forehead rest and replace the chin
rest paper.
4Have the patient sit down.
5Align the height of the patient’s eye.
6Move the stage position.
7Align the retinal camera to the eye to be photographed.
8Adjust brightness in the field of view for observation.