5.4 Mydriatic Mode

The Mydriatic mode is designed for photographing the
patient’s eye that a mydriatic is applied. To switch from the
non-mydriatic mode to the mydriatic mode, hold down the
observation mode switch for approx. 2 or more seconds.
The observation mode lamp Myd turns on.
The viewfinder is used for retinal observation.
Check that the diopter of the viewfinder is aligned (see
page 12).

5.4.1 Color Photography Mode

1Enter the study information into the computer.
Enter the study information with the control software.
2Select the color photography mode.
Press the COLOR switch.
Or select the photography mode with the control
software.
The COLOR lamp turns on.
3Disinfect the forehead rest and replace the chin rest paper.
If chin rest paper is not used, disinfect the chin rest.
4Have the patient sit down.
Have the patient place his or her chin against the chin rest and his or her forehead against the
forehead rest.
CAUTION
To prevent the risk of infection, wipe the forehead rest with a disinfectant
such as ethanol or glutaraldehyde whenever the patient is changed.
For details on the disinfection procedure, consult a specialist.
CAUTION To ensure cleanliness, replace the chin rest paper whenever the patient is
changed.
!
!