5. Basic Photography
-30-
5.4.2 Red-free Photography Mode
1Enter the study information into the computer.
2Select the red-free photography mode.
Press the RED FREE switch. Or select the photography mode with the control software.
The RED FREE lamp turns on.
3Perform the procedure starting from step 3 in 5.4.1 Color Photography Mode (see page24).
5.4.3 Cobalt Photography Mode
1Enter the study information into the computer.
2Select the cobalt photography mode.
Press the COBALT switch. Or select the photography mode with the control software.
The COBALT lamp turns on.
3Perform the procedure starting from step 3 in 5.4.1 Color Photography Mode (see page24).
5.4.4 Fluorescein Angiography Mode
1Enter the study information into the computer.
2Select the fluorescein angiography mode.
Press the FLUO switch. Or select the photography mode with the control software.
The FLUO lamp turns on.
3Disinfect the forehead rest and replace the chin rest papar.
4Have the patient sit down.
5Align the height of the patient’s eye.
6Move the stage position.
7Align the camera to the eye to be photographed.
8Adjust brightness in the field of view for observation.
9Determine the area to be photographed.
Concerning the barrier filter
During retinal observations in the fluorescein angiography mode, each time the BA switch is
pressed, the barrier filter is inserted or disengaged. However, while images are being taken, the
filter is inserted automatically.
While the barrier filter is inserted, the lamp above the button lights brightly.