5. Basic Photography
-32-

5.4.5 Fundus Autofluorescence Mode

1Enter the study information into the computer.
2Select the fundus autofluorescence mode.
Press the FAF switch.
Or select the photography mode with the control
software.
The FAF lamp turns on.
3Perform the procedure starting from step 3 in 5.4.1 Color Photography Mode (see page24).
Concerning the adjustment of the flash intensity (for fundus autofluorescence mode)
The flash intensity is already set to the standard level for each photography mode.
To adjust the intensity, press the flash intensity switches.
Each time one of these switches is pressed, the intensity is increased or reduced by steps of 0.3.
The flash intensity selected is shown by the lighted area of the flash intensity indicator.
Note: For details on the flash intensity adjustment in the color, red free, cobalt
photography mode, and fluorescein angiography mode, refer to Concerning the
adjustment of the flash intensity (for color, red free, cobalt photography mode, and
fluorescein angiography mode) (page28).
Standard
Reduced by 19% (equivalent to a step of 0.3)
of the standard
Reduced by 34% (equivalent to a step of 0.6)
of the standard
Reduced by 47% (equivalent to a step of 0.9)
of the standard
Reduced by 56% (equivalent to a step of 1.2)
of the standard
Reduced by 65% (equivalent to a step of 1.5)
of the standard