7. Daily Inspection and Maintenance
-48-

7.5 Removing and Installing the Digital Camera

It is not necessary to remove and re-install the digital camera during normal usage.
The procedure below is used to remove and install the digital camera during maintenance and inspection of
the digital camera or other work.
Note: Do not touch the lens part of the main unit and the mirror part of the digital camera when
attaching and detaching the digital camera from the retinal camera body. If any dirt,
fingerprints, dust, and other foreign objects attaches to the lens or mirror part, you cannot
take a good image.

7.5.1 Removing the Digital Camera

Note: If the digital camera is removed an d left unattached, dust and other foreign objects can
enter the main unit and digital camera. After removing the digital camera, be sure to also
attach the caps to each mount.
1Turn off the power for the retinal camera and digital camera.
Set the power switch of the digital camera to the OFF position.
Press the power switch of the retinal camera to the O side (OFF).
2Remove the camera cover.
Loosen the screws of the camera cover to remove the
cover.
3Disconnect the cables.
Remove the respective cables from the digital terminal,
PC terminal, and remote control terminal.