Dip Switch Setup

To set up the KT6 for your specific application, you might need to change some of the Dip Switches. The switches are located under the chart in the lower half of the dial plate. A pointed object can be used to change the settings. Remember to install the correct chart to match corresponding dip switch setting.

Slide toggle to this side for ON

All switches are

1

2

3

4

5

6

7

8

shown in OFF

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

position.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Slide toggle to this side for OFF

 

 

KT6 has two different recording time options

1 day

#3 ON

7 day

#3 OFF

˚F/˚C

 

˚C

#4 ON

˚F

#4 OFF

Temperature Range

 

0 to 100˚

#5 OFF

 

#6 OFF

 

#7 OFF

0 to 50˚

#5 ON

 

#6 OFF

 

#7 OFF

50 to 100˚

#5 OFF

 

#6 ON

Calibration

Your instrument was carefully tested and calibrated before being shipped from the factory. Additional calibration is not required. However, should calibration be desired in the future, we recommend that it be at our lab. Call Customer Service at (630) 543-3747. If you wish to calibrate yourself, follow these procedures:

1.To activate the calibration mode, turn the unit OFF. Now press the “ON/OFF” key and the “ALARM” key at the same time. The unit is in the “User Calibration” mode when the red LED is ON. The display will show a “C” during the warm up period and then will flash between 0 and F or 0 and C (depending on measurement selected). Match the reading

of the recorder to the reading of the reference standard.

Rev. 03/04

 

#7 OFF

0 to 250˚

#5 ON

 

#6 ON

 

#7 OFF

0 to 500˚F (only ˚F)

#5 OFF

 

#6 OFF

 

#7 ON

-50 to 50

#5 ON

 

#6 OFF

 

#7 ON

0 to -50

#5 OFF

 

#6 ON

 

#7 ON

0 to -100

#5 ON

 

#6 ON

 

#7 ON

ALARM SPEAKER:

The alarm speaker may be activated or deactivated with the use of Dip Switch #8

Deactivated

#8

ON

Activated

#8

OFF

The Alarm Speaker will sound during minimum, maximum or probe missing alarm conditions.

NOTE: dip switch settings are only updated when the unit is turned on.

2.To raise the unit of measurement, press the “CHART ROTATION (UP ARROW)” key. To lower the unit of measurement, press the “HOME (DOWN ARROW)” key. To ensure that you have matched the standard, allow the pens to stabilize for at least 30 seconds before exiting calibration mode. If your KT6 only has one pen, skip to Step 4.

3.Press the “ALARM” button for 2 seconds to switch from Pen 1 to Pen 2. Repeat step 2 for Pen 2 temperature calibration.

4.When calibration is complete, simply press the “ON/OFF” key to save the calibration settings. Calibration is stored in memory even after you turn the unit OFF. User Calibration information will not be lost if AC Power fails.

DICKSON

 

 

Specifications

Product

Display Symbols

Operating Information &

 

Accessories

Charts &

 

Anatomy

Instrument

Installation

Started/

Getting

 

Operation

Keypad

Calibration

Setup &

Dip Switch

Returns

Warranty &

Troubleshooting

Page 8
Image 8
Dickson Industrial KT6 manual Dip Switch Setup, Calibration, ˚F/˚C, Temperature Range, Alarm Speaker