Dickson Industrial KT6 manual Troubleshooting, Warranty, Factory Service & Returns

Models: KT6

1 10
Download 10 pages 4.65 Kb
Page 9
Image 9

Troubleshooting

Symptom

Cause

Check/Remedy

Pen trace too fine or absent

Tip too sharp, cartridge

Sand tip, moisten or replace pen.

Readings of pen and display do

Wrong chart, pen not on

Check Dip Switch #4, 5, 6.

properly, line thickness

Push pen completely onto pen arm.

not match

 

Check to see that proper chart is

 

 

installed

UND

Below range of recorder

Check temperature range.

Out of calibration

Harsh environments, stressful

See Calibration Procedure on page 8

or

conditions, or time.

or

Questionable accuracy

 

Return for factory re-calibration

OVR

Above range of recorder

Check temperature range

PRB

Sensor missing

Connect transducer to back of

 

the recorder (see diagram on page 3)

 

 

Warranty

 

 

Dickson warrants that the products it sells will be free from defects in material and workmanship under normal use and service for a period of twelve months after delivery. In the event of a claim under this warranty, the product or part must be returned to the factory for repair or replacement (shipping pre-paid) with a Return Authorization Number (see Return Information above). It will be repaired at Dickson’s option without charge. This warranty DOES NOT cover routine calibration, pen, chart and battery replacement. The foregoing warranty and remedy are exclusive and in lieu of all other warranties either expressed or implied. Dickson shall not be liable for consequential or incidental damages resulting from failure or malfunction of its products. Dickson makes no warranty for products not manufactured by it or for any products modified by buyer, or subject to misuse or neglect.

Factory Service & Returns

Contact the factory (630-543-3747) for a Return Authorization (RA) Number before returning any instrument. The model number, serial number and a purchase order number will be requested before an RA number is issued.

Carefully repack the instrument, label the outside of the box with the RA# and return the instrument (freight pre-paid) to Dickson.

All instruments that do not have the RA# clearly marked on the outside of the box will be refused. When returning instruments for credit, please include all accessories in shipment.

Calibration/Freight charges are non-refundable.

NOTE: Dickson shall not be liable for consequential or incidental damages resulting from failure or malfunction of its products.

Customer Satisfaction

Dickson takes pride in providing you, the customer, with the highest quality instrumentation. We welcome the opportunity to help you in any way possible. Whether it be a question or a new idea in documentation, the Dickson Company would like to hear your response. Please call our Customer Service Department at 1-800-323-2448 or (630) 543-3747 (in Illinois).

DICKSON

930 South Westwood Avenue

Addison, Illinois 60101

Phone: (630) 543-3747E-mail: DicksonCSR@dicksonweb.com

Rev. 03/04

DICKSON

 

 

Specifications

Product

Display Symbols

Operating Information &

 

Accessories

Charts &

 

Anatomy

Instrument

Installation

Started/

Getting

 

Operation

Keypad

Calibration

Setup &

Dip Switch

Returns

Warranty &

Troubleshooting

Page 9
Image 9
Dickson Industrial KT6 manual Troubleshooting, Warranty, Factory Service & Returns