Guidance and manufacturer’s declaration-electromagnetic immunity for all

EQUIPMENT AND SYSTEMS

Guidance and manufacturer’s declaration-electromagnetic emissions

The 2401LM Touchmonitor for Healthcare Applications is intended for use in the electromagnetic environment specified below. The customer or the user of the 2401LM Touchmonitor for Healthcare Applications should assure that it is used in such an environment.

Emissions test

Compliance

Electromagnetic environment-guidelines

 

 

 

RF emissions

Group 1

The 2401LM Touchmonitor for Healthcare Applications uses RF

CISPR 11

 

energy only for its internal function. Therefore, its RF emissions

 

 

are very low and are not likely to cause any interference in nearby

 

 

electronic equipment.

 

 

 

RF emissions

Class A

The 2401LM Touchmonitor for Healthcare Applications is suitable

CISPR 11

 

for use in all establishments other than domestic and those directly

 

 

connected to a low voltage power supply network which supplies

 

 

buildings used for domestic purposes.

Harmonics

Class D

 

emissions

 

 

IEC 61000-3-2

 

 

 

 

 

Voltage

Complies

 

fluctuations/flicker

 

 

emissions

 

 

IEC 61000-3-3

 

 

 

 

 

 

User Manual 2401LM

© 2013 Elo Touch Solutions, Inc. All rights reserved.

SW200127 Rev A - Page 7 of 32

Page 7
Image 7
Elo TouchSystems 2401LM user manual Equipment and Systems