Guidance and manufacturer’s declaration-electromagnetic immunity for all EQUIPMENT AND

SYSTEMS

Guidance and manufacturer’s declaration-electromagnetic immunity

The 2401LM Touchmonitor for Healthcare Applications is intended for use in the electromagnetic environment specified below. The customer, or the user of the 2401LM Touchmonitor for Healthcare Applications, should assure that it is used in such an environment.

Immunity Test

IEC 60601

Compliance Level

Level

 

 

 

 

 

Electrostatic

± 6 kV contact

± 6 kV contact

Discharge (ESD)

± 8 kV air

± 8 kV air

IEC 61000-4-2

 

 

 

 

 

Electrical Fast

± 2 kV for power

± 2 kV for power

Transient/Burst

supply lines

supply lines

IEC 61000-4-4

± 1 kV for

± 1 kV for input/output

 

input/output lines

Lines

 

 

 

Surge

± 1 kV line(s) to

± 1 kV line(s) to

IEC 61000-4-5

line(s)

line(s)

 

± 2 kV line(s) to

± 2 kV line(s) to earth

 

earth

 

 

 

 

Voltage Dips, Short

<5% UT

<5% UT

Interruption and

(>95% dip in UT)

(>95% dip in UT)

Voltage Variations

for 0.5 cycle

for 0.5 cycle

on Power Supply

40% UT

40% UT

Input Lines

 

(60% dip in UT)

(60 % dip in UT)

IEC 61000-4-11

for 5 cycles

for 5 cycles

70% UT

70% UT

 

 

(30% dip in UT)

(30% dip in UT)

 

for 25 cycles

for 25 cycles

 

<5% UT

<5% UT

 

(>95% dip in UT)

(>95% dip in UT)

 

for 250 cycles

for 250 cycles

 

 

 

Power Frequency

3 A/m

3 A/m

(50/60 Hz)

 

 

Magnetic Field

 

 

IEC 61000-4-8

 

 

 

 

 

Electromagnetic Environment - Guidelines

Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.

Mains power quality should be that of a typical commercial or hospital environment.

Mains power quality should be that of a typical commercial or hospital environment.

Mains power quality should be that of a typical commercial or hospital environment. If the user of the 2401LM Touchmonitor for Healthcare Applications requires continued operation during power mains interruptions, It is recommended that the 2401LM Touchmonitor for Healthcare Applications be powered from an uninterruptible power supply or a battery.

Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.

NOTE UT is the A.C. mains voltage prior to application of the test level.

 