Emerson 4.2396P and 396PVP ORP Sensors, 4.2.1Sensor preparation, 4.2.2ORP Calibration, August

Models: 396P 396PVP

1 54
Download 54 pages 50.65 Kb
Page 42
Image 42

Section 4: Start-Up & Calibration

396P + 396PVP Sensors Instruction Manual

August 2013

LIQ_MAN_396P_396PVP

 

 

2.Measure and record the pH of the process solution sample with another temperature compen- sated, calibrated pH instrument. For best results, standardization should be performed at the process temperature.

3.Adjust the analyzer/transmitter value to the standardized value.

4.2396P and 396PVP ORP Sensors

4.2.1Sensor preparation

Most industrial applications have a number of ORP reactions occurring in sequence or simultane- ously. There can be several components that are oxidized or reduced by the reagents that are used. Theoretically, the ORP potential is absolute because it is the result of the oxidation-reduc- tion equilibrium. However, the actual measured potential is dependent on many factors, including the condition of the surface of the ORP platinum electrode. Therefore, the sensor should be allowed 1-2 hours to become “conditioned” to the stream when first set-up or after being cleaned.

4.2.2ORP Calibration

1.Make a temporary electrical connection between the sensor and the instrument.

2.Obtain an ORP standard solution, or a standard solution can also be made quite simply by adding a few crystals of quinhydrone to either pH 4 or pH 7 buffer. Quinhydrone is only slightly soluble therefore a few crystals will be required. (Refer to Section 4.3. for an alternate ORP standard solution).

3.Immerse the sensor in the standard solution. Allow 1-2 minutes for the ORP sensor to stabilize.

4.Adjust the standardize control of the instrument to the solution value shown in Table 5-1 (below) or on the label of the standard solution. The resulting potentials, measured with a clean platinum electrode and saturated KCl/AgCl reference electrode, should be within ±20 millivolts of the value. Solution temperature must be noted to ensure accurate interpretation of results. The ORP value of saturated quinhydrone solution is not stable over long periods of time. Therefore, these standards should be made up fresh each time they are used.

5.Remove the sensor from the buffer, rinse and install in the process.

TABLE 4-1. ORP of Saturated Quinhydrone Solution (In Millivolts)

 

pH 4 Solution

pH 7 Solution

Temp °C

20

25

30

20

25

30

Millivolt Potential

268

264

260

94

87

80

36

Start-Up and Calibration

Page 42
Image 42
Emerson 4.2396P and 396PVP ORP Sensors, 4.2.1Sensor preparation, 4.2.2ORP Calibration, Start-Up& Calibration, August