Emerson 396PVP Diagnostics and Troubleshooting, Diagnostics & Troubleshooting

Models: 396P 396PVP

1 54
Download 54 pages 50.65 Kb
Page 47
Image 47

396P + 396PVP Sensors Instruction Manual

Section 6: Diagnostics & Troubleshooting

LIQ_MAN_396P_396PVP

August 2013

 

 

Section 6: Diagnostics and Troubleshooting

6.154e/56/1056/1057/1066/3081/4081/5081/XMT Diagnostics and Troubleshooting

Many Rosemount Analytical Instruments and Transmitters automatically search for fault condi- tions that would cause an error in the measured pH value. Refer to the applicable Instruction Manual for a complete description of the analyzer’s fault conditions.

Table 7-1, below, lists some of the diagnostic messages that indicate a possible sensor problem. A more complete description of the problem and a suggested remedy corresponding to each message is also listed.

TABLE 6-1. Troubleshooting with Advanced Diagnostics

DIAGNOSTIC MESSAGE

DESCRIPTION OF PROBLEM

REMEDY

 

 

 

“Calibration Warning”

1. Aged glass.

1. Perform buffer calibration.

CALIbrAtE

2. Sensor not immersed.

2. Be sure electrode measuring tip

 

 

is in process.

 

 

 

“Cracked glass failure”

Broken or cracked glass.

Replace sensor.

6LASS fAIL

 

 

 

 

 

“High reference impede”

1. Liquid junction coated.

1. Clean sensor; replace if necessary.

rEF fAIL or rEF WjArn

2. Reference Cell gel depleted.

2. Replace sensor.

 

3. Sensor not immersed.

3. Be sure electrode measuring tip

 

 

is in process.

 

 

 

“Input voltage high”

pH input shorted or sensor

Check wiring. Replace sensor if

“Input voltage low”

miswired.

necessary.

 

 

 

“Old glass warning”

1. Glass electrode worn out.

1. Replace sensor.

6LaSS WjArn

2. Sensor not immersed.

2. Be sure electrode measuring tip

 

 

 

is inprocess.

 

 

 

“Reference offset err”

Reference electrode poisoned.

Replace sensor.

(offline only)

 

 

Std Err

 

 

 

 

 

“Ref voltage high”

1. Reference shorted or sensor miswired.

Check wiring and installation.

“Ref voltage low”

2. Sensor not immersed

Replace sensor if necessary.

“Sensor line open”

1. Open wire between sensor and analyzer.

1. Check sensor wiring.

LInE FAIL

2. Interconnecting cable greater than

2. Relocate analyzer.

 

1000 ft.

 

 

 

 

“Sensor miswired”

1. Open wire between sensor and analyzer.

1. Check wiring.

 

2. Bad preamplifier.

2. Replace preamplifier. (Code 02 only)

 

 

 

“Temp error high”

1. Open or shorted RTD.

1. Replace sensor.

“Temp error low”

2. Temperature out of range.

2. Check process temperature.

tEMP HI

 

 

tEMP LO

 

 

 

 

 

Diagnostics and Troubleshooting

41

Page 47
Image 47
Emerson Diagnostics and Troubleshooting, Diagnostics & Troubleshooting, 396P + 396PVP Sensors Instruction Manual