MODEL 396R pH/ORP | SECTION 6.0 |
| MAINTENANCE |
4.Following the caution above, wash the glass bulb in dilute 5% hydrochloric acid solution and then rinse it thoroughly in tap water. Replace the sen- sor if it cannot be cleaned. If the glass bulb appears clean, proceed to step 5.
5.Buffer calibrate the sensor (Refer to Section 5.0). If the sensor appears to respond sluggishly to pH change, soaking it overnight in a weak acid solution (5% hydrochloric acid) may improve its response. Be sure to follow the CAUTION above and to rinse the sensor’s tip thoroughly with tap water. If the sensor will not calibrate, it must be replaced.
6.4Cleaning Platinum Electrode. The electrode is never exposed to these undesirable compounds. In the event poisoning is suspected, the electrode can be restored to normal operation by simply cleaning the plat- inum electrode with baking soda. Polish it by rubbing it with a damp paper towel and baking soda until a bright, shiny appearance is attained.
6.5Automatic Temperature Compensator. The tem- perature compensator element is temperature sensi- tive and can be checked with an ohmmeter. Resistance increases with temperature.
The 3K element will read 3000 ohms ± 1% at 25°C (77°F) and a
RT=Ro [l+R1
Where RT = Resistance
T = Temperature in °C
Refer to Table
TABLE
Ro and R1 VALUES FOR TEMPERATURE
COMPENSATION ELEMENTS
Temperature | Ro | R1 | |
Compensation Element | |||
|
| ||
3K | 2934 | .0045 | |
107.7 | .00385 | ||
|
|
|
TABLE
TEMPERATURE vs RESISTANCE OF AUTO
T.C. ELEMENTS
| Resistance | |
Temperature °C | (Ohms) ±1% | |
| 3K |
|
0 | 2670 | 100.0 |
10 | 2802 | 103.8 |
20 | 2934 | 107.7 |
25 | 3000 | 109.6 |
30 | 3066 | 111.5 |
40 | 3198 | 115.4 |
50 | 3330 | 119.2 |
60 | 3462 | 123.1 |
70 | 3594 | 126.9 |
80 | 3726 | 130.8 |
90 | 3858 | 134.6 |
100 | 3990 | 138.5 |
|
|
|
6.6Sensor Tube Replacement When Used With A Sensor Head Junction Box. Replacement of the retraction versions sensor tube assembly involves the removal and installation of two sets of male connectors: One at the process end of the sensor, and the other at the junction box end (See Figures
1.Remove sensor from process before proceeding. The junction box with attached male connector must be recovered from the old sensor for reuse. Unscrew the junction box cover and set aside. Disconnect electrical connections from printed cir- cuit board inside junction box. Disconnect BNC connector to preamp. Unscrew hex nut (D) from male connector body (C). Separate junction box from used sensor. Set aside.
2.Pry off split ferrule from sensor and set aside for reuse. Remove hex nut (D) and set aside for reuse. Check that the internal
3.Remove hex nut (B) from male connector body (A) at process end of sensor and set aside. Slide the Teflon ferrule and the male connector off sensor in the direction of junction box and set
NOTE
If stainless steel ferrule was used, male connector body (A) will have to be dis- carded with the sensor tube.
33