Emerson 396RVP Section Diagnostics and Troubleshooting, Troubleshooting with Diagnostics

Models: 396R 396RVP

1 44
Download 44 pages 57.38 Kb
Page 40
Image 40

MODEL 396R pH/ORP

SECTION 7.0

 

DIAGNOSTICS AND TROUBLESHOOTING

SECTION 7.0

DIAGNOSTICS AND TROUBLESHOOTING

7.1DIAGNOSTICS AND TROUBLESHOOTING WITH MODEL 54/81/3081 pH/ORP. The Model 54 and 54e Analyzers and Models 81 and 3081 pH Transmitters automatically search for fault conditions that would cause an error in the measured pH value, as does the Model 1054A/B pH/ORP Analyzer to a lesser degree. Refer to the respective manual for a complete description of the analyzer’s fault conditions.

Table 7-1 lists the Analyzer/Transmitters diagnostic messages that indicate a possible sensor problem. A more complete description of the problem and a suggested remedy corresponding to each message is also listed.

TABLE 7-1 Troubleshooting with Diagnostics

DIAGNOSTIC MESSAGE

 

 

 

 

54 and 54e

 

 

 

 

81/3081

DESCRIPTION OF PROBLEM

REMEDY

 

 

 

 

“Calibration Warning”

1. Aged glass.

1.

Perform buffer calibration.

 

2.

Sensor not immersed.

2.

Be sure electrode measuring tip is in

DBMJcsBtF

 

 

 

process.

“Cracked glass failure”

Broken or cracked glass.

Replace Sensor.

7MBTT!GBJM

 

 

 

 

“High reference imped”

1.

Liquid junction coated.

1.

Clean sensor; replace if necessary.

 

2.

Reference Cell gel depleted.

2.

Replace sensor.

sFG!GBJM or sFG!WBso

3.

Sensor not immersed.

3.

Be sure electrode measuring tip is in

 

 

 

 

process.

“Input voltage high”

pH input shorted or sensor.

Check wiring. Replace sensor if

“Input voltage low”

miswired.

necessary.

 

 

 

 

 

“Old glass warning”

1.

Glass electrode worn out.

1.

Replace sensor.

7MbTT!WBso

2.

Sensor not immersed.

2.

Be sure electrode measuring tip is in

 

 

 

process.

 

 

 

 

“Reference offset err”

Reference electrode poisoned.

Replace sensor.

(offline only)

 

 

 

 

Tte!Fss

 

 

 

 

“Ref voltage high”

1.

Reference shorted or sensor

Check wiring. Replace sensor if

“Ref voltage low”

 

miswired.

necessary.

 

2.

Sensor not immersed.

 

 

“Sensor line open”

1.

Open wire between sensor and analyzer.

1.

Check sensor wiring.

 

2.

Interconnecting cable greater than

2.

Relocate analyzer.

MJoF!GBJM

 

1000 ft.

 

 

 

 

 

 

 

“Sensor miswired”

1.

Open wire between sensor and analyzer.

1.

Check wiring.

 

2.

Bad preamplifier.

2.

Replace preamplifier.

 

 

 

 

“Temp error high”

1.

Open or shorted RTD.

1. Replace sensor.

“Temp error low”

2. Temperature out of range.

2. Check process temperature.

tFMQ!IJ

 

 

 

 

tFMQ!MP

 

 

 

 

36

Page 40
Image 40
Emerson 396R Section Diagnostics and Troubleshooting, Troubleshooting with Diagnostics, Diagnostic Message, 54e 81/3081