MODEL 5081T-FF/FI

MAINTENANCE

MAINTENANCE

OVERVIEW

This section gives general procedures for routine maintenance of the 5081-T transmitter. The transmitter needs almost no routine maintenance.

TRANSMITTER MAINTENANCE

Periodically clean the transmitter window with household ammonia or glass cleaner. The detector for the infrared remote controller is located behind the window at the top of the transmitter face. The window in front of the detector must be kept clean.

Most components of the transmitter are replaceable. Refer to figure below and table on next page for parts and part numbers.

Exploded View of Model 5081-T-FF/FI Transmitter

Three screws (part 13 in the drawing) hold the three circuit boards in place. Removing the screws allows the display board (part 2) and the CPU board (part 3) to be easily removed. A ribbon cable connects the boards. The cable plugs into the CPU board and is permanently attached to the display board. A 16 pin and socket connector holds the CPU and analog (part 4) boards together. Five screws hold the terminal block (part 5) to the center housing (part 7), and the 16 pins on the terminal block mate with 16 sockets on the back side of the analog board. Use caution when separating the terminal block from the analog board. The pin and socket connection is tight.

21

Page 21
Image 21
Emerson 5081T-FF/FI Overview, Transmitter Maintenance, Exploded View of Model 5081-T-FF/FI Transmitter