Analog Output Setup:
Press the ENTER key or the →
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| CLD |
|
|
|
|
|
|
|
| 95.0 ppm |
|
|
|
|
|
|
| |||
|
|
|
| - ANALOG OUTPUT SETUP - |
|
|
|
|
|
|
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||
| Output number: |
|
|
|
| 1 |
|
|
|
|
|
|
|
| ||||||
| Choose signal source module... |
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||
| Choose signal... |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||
| Signal value for 0% output: |
|
|
| 0.00 ppm |
|
|
|
|
|
|
| ||||||||
| Signal value for 100% output: |
| 2500.00 ppm |
|
|
|
|
|
|
| ||||||||||
| Output current range: |
|
|
|
| 0...20 mA |
|
|
|
|
|
|
| |||||||
| Hold output during calibration? |
|
|
|
| No |
|
|
|
|
|
|
| |||||||
| Signal name: |
|
|
|
|
| NO/NOx |
|
|
|
|
|
|
| ||||||
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||||
| Current signal value: |
|
|
|
| 95.00 ppm |
|
|
|
|
|
|
| |||||||
| Source module: |
|
|
|
|
|
| CLD |
|
|
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| EXPERT |
|
|
|
|
| ADJUST |
| INFO |
|
|
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||
| F1 | F2 | F3 | F4 |
| F5 |
| |||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Configuration of the analog output:
1) Select the analog output number:
You can set up all parameters in the menu "ANALOG OUTPUT SETUP" separately for each of the available analog outputs. The number of outputs depends on the hardware configuration of the SIO module (minimum: 2, maximum: 8).
¨Press the ENTER key or the →
¨Confirm the number selected with the ENTER key.
2) Choose the signal source module:
¨Change with the ↓
¨Change to the submenu "ANALYZER MODULES" with the ENTER key or the →
CLD | 95.0 ppm |
- ANALYZER MODULES -
CLD
MENU BACK
>>>
<<<
F1 | F2 | F3 | F4 | F5 |
90003749(1)
NGA 2000 | 5 - 31 |