Analog Output Setup:

Press the ENTER key or the -key in the line "Analog output setup..." to change to the corresponding submenu:

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CLD

 

 

 

 

 

 

 

 

95.0 ppm

 

 

 

 

 

 

 

 

 

 

 

- ANALOG OUTPUT SETUP -

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Output number:

 

 

 

 

1

 

 

 

 

 

 

 

 

 

Choose signal source module...

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Choose signal...

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Signal value for 0% output:

 

 

 

0.00 ppm

 

 

 

 

 

 

 

 

Signal value for 100% output:

 

2500.00 ppm

 

 

 

 

 

 

 

 

Output current range:

 

 

 

 

0...20 mA

 

 

 

 

 

 

 

 

Hold output during calibration?

 

 

 

 

No

 

 

 

 

 

 

 

 

Signal name:

 

 

 

 

 

NO/NOx

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Current signal value:

 

 

 

 

95.00 ppm

 

 

 

 

 

 

 

 

Source module:

 

 

 

 

 

 

CLD

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

EXPERT

 

 

 

 

 

ADJUST

 

INFO

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

F1

F2

F3

F4

 

F5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Configuration of the analog output:

1) Select the analog output number:

You can set up all parameters in the menu "ANALOG OUTPUT SETUP" separately for each of the available analog outputs. The number of outputs depends on the hardware configuration of the SIO module (minimum: 2, maximum: 8).

¨Press the ENTER key or the -key to select the output number and adjust the value you want with the -key or the -key.

¨Confirm the number selected with the ENTER key.

2) Choose the signal source module:

¨Change with the -key to the line "Choose signal source module..."

¨Change to the submenu "ANALYZER MODULES" with the ENTER key or the -key. The tags of all analyzer modules combined with the platform or the MLT analyzer will appear:

CLD

95.0 ppm

- ANALYZER MODULES -

CLD

MENU BACK >>> <<<

F1

F2

F3

F4

F5

90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98

NGA 2000

5 - 31