Serial interface setup:
Press the ENTER key or the →
CLD |
|
|
| 95.0 ppm |
| - SERIAL INTERFACE SETUP - |
| ||
Baudrate: |
|
|
| 19200 |
Data bits: |
|
|
| 8 |
Stop bits: |
|
|
| 1 |
Parity: |
|
|
| None |
Echo mode: |
|
|
| Disabled |
Handshake: |
|
|
| Xon/Xoff |
Type of installed serial interface: |
| RS232 | ||
EXPERT |
|
|
| INFO |
F1 | F2 | F3 | F4 | F5 |
In the submenu "SERIAL INTERFACE SETUP" you can determine the parameters of the data transfer between the MLT analyzer or platform and external implements. The setups in this menu are depending on the configuration of the analyzer resp. platform and the corresponding implement. The specification of the serial interface is described in another manual.
Set up parameters:
¨Select any line of variables with the ↓
¨Select the variable with the ENTER key or the →
¨Select any digit with the ←
resp. select the whole parameter with the −
¨Confirm the new value with the ENTER key or cancel and go back to the last value with the F2 key.
Options: |
|
|
|
|
|
|
Baudrate: | 300 | 1200 | 2400 | 4800 | 9600 | 19200 |
Data bits: | 7 | 8 |
|
|
|
|
Stop bits | 1 | 2 |
|
|
|
|
Parity: | None | Even | Odd |
|
|
|
Echo mode: | Enabled | Disabled |
|
|
|
|
Handshake: | None | Xon/Xoff |
|
|
|
|
Type of installed |
|
|
|
|
|
|
serial interface: | RS 232 | RS 485_2X | RS 485_4X | None |
|
|
90003749(1)
NGA 2000 | 5 - 35 |