7) Fine adjustment of the analog output:

If you press the F4 key in the menu "ANALOG OUTPUT SETUP" you can change to the submenu "OUTPUT FINE ADJUSTMENT", where you can do the fine adjustments of the analog output:

CLD

 

 

 

95.0 ppm

 

- OUTPUT FINE ADJUSTMENT -

 

Output number:

 

 

1

Operation mode:

 

 

Normal

Fine adjustment for 0% output:

 

4097

Fine adjustment for 100% output:

 

799

EXPERT

 

 

BACK

INFO

F1

F2

F3

F4

F5

In the line "Output number" the output number selected in the menu "ANALOG OUTPUT SETUP" will appear. If you want you can change the number in the submenu "OUTPUT FINE ADJUSTMENT":

¨Press the ENTER key or the -key to select the output number and adjust the value you want with the -key or the -key. Options: 1, ..., 8.

¨Confirm the number selected with the ENTER key.

Set up fine adjustment of the analog output:

¨Press the ENTER key or the -key in the line "Operation mode" to select the parameter and adjust one of the following parameters with the -key or the -key:

·Normal: The absolute measurement signal will be sent to the analog output.

·Adjust 0V: Coordination between the display and the analog output for 0V with fine adjustment 0%.

·Adjust 10V: Coordination between the display and the analog output for 10V with fine adjustment 100%.

¨Confirm the new parameter with the ENTER key.

¨Press the ENTER key or the -key in the line "Fine adjustments for 0% output" or "Fine adjustment for 100% output" to select the corresponding value.

¨Adjust the value you want with the -key or the -key and confirm it with the ENTER key. Options: 3500 - 4800 for 0% and 750 - 900 for 100%.

The last three lines of variables in the menu "Analog output setup" only display some values of the analog output! They are not available.

5 - 34

NGA 2000

90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98