7) Fine adjustment of the analog output:
If you press the F4 key in the menu "ANALOG OUTPUT SETUP" you can change to the submenu "OUTPUT FINE ADJUSTMENT", where you can do the fine adjustments of the analog output:
CLD |
|
|
| 95.0 ppm |
| - OUTPUT FINE ADJUSTMENT - |
| ||
Output number: |
|
| 1 | |
Operation mode: |
|
| Normal | |
Fine adjustment for 0% output: |
| 4097 | ||
Fine adjustment for 100% output: |
| 799 | ||
EXPERT |
|
| BACK | INFO |
F1 | F2 | F3 | F4 | F5 |
In the line "Output number" the output number selected in the menu "ANALOG OUTPUT SETUP" will appear. If you want you can change the number in the submenu "OUTPUT FINE ADJUSTMENT":
¨Press the ENTER key or the →
¨Confirm the number selected with the ENTER key.
Set up fine adjustment of the analog output:
¨Press the ENTER key or the →
·Normal: The absolute measurement signal will be sent to the analog output.
·Adjust 0V: Coordination between the display and the analog output for 0V with fine adjustment 0%.
·Adjust 10V: Coordination between the display and the analog output for 10V with fine adjustment 100%.
¨Confirm the new parameter with the ENTER key.
¨Press the ENTER key or the →
¨Adjust the value you want with the −
The last three lines of variables in the menu "Analog output setup" only display some values of the analog output! They are not available.