¨Go to the line you want with the −
¨Select the tag of the channel with the ENTER key or the →
The display will go back to the menu "ANALOG OUTPUT SETUP" automatically. The tag of the channel selected will appear in the line "Source module".
3) Select any signal:
¨Change to the line "Choose signal" with the ↓
¨Press the ENTER key or the →
CLD |
|
|
| 95.0 ppm |
|
| - SIGNALS - |
|
|
|
|
|
| Sample flow: |
|
|
| Sample press.: | |
|
|
|
| Ozonator: |
|
|
| Converter temp.: | |
|
|
| Block temperature: | |
|
|
| Detector temperature: | |
|
|
|
| NO/NOx: |
|
|
|
| Noise level: |
MENU | BACK | >>> | <<< | INFO |
F1 | F2 | F3 | F4 | F5 |
¨Change to the line you want with the −
You can change with the F3 key (>>>) to further menu pages containing other available signals:
·t90 time
·Output delay time
·Calibration status
·Linearizer
·Measurement mode
·Operational state
·Health
·Interference
·Validity
·Raw signal
·Primary Variable
¨Press the ENTER key or the →
The display will go back to the menu "ANALOG OUTPUT SETUP" automatically. The signal selected will appear in the line "Signal name".
4) Determination of the output signal values:
You can determine the signal value for the 0% output and the 100% output in the lines "Signal value for 0% output" or "Signal value for 100% output". So you have the possibility to zoom out a certain part of the whole range.
5 - 32 | NGA 2000 |