5.4.2 System Set Up

Display resolution

Main Menu — Expert controls and set up

 

 

 

 

 

 

System set up

 

 

 

 

 

 

 

Display resolution

 

 

 

 

 

CLD

 

 

 

95.0 ppm

 

 

Display resolution

 

Maximum total number of digits:

 

4

Digits after the decimal point:

 

2

HOME

ESCAPE

 

INFO

F1

F2

F3

F4

F5

In the menu "Display resolution" you can determine the number of digits for the display of the measured values. These numbers do not alter the accuracy of measurement. The adjustments are not valid for the analyzer module!

Set up parameters:

¨Select one of the two lines of variables with the -key or the -key.

¨Select the number with the ENTER key or the -key.

¨Select the digit with the -key or the -key. Options: Maximum total number of digits: 3, 4, 5, 6

Digits after the decimal point: 0, 1, 2, 3.

¨Confirm the new value with the ENTER key or cancel and go back to the last value with the F2 key.

Example:

¨Measured number: 1.2345

¨Select 3 for total number of digits and 2 for digits after decimal point.

¨The screen will display: 1.23

5 - 50

NGA 2000