II) V24/RS232/485 Interface - Single Analyzers and Systems

ELIN – Write command "Linearization values (X/Y)"

With this write command the analyzers in a system or the single analyzer will get the linearization values for the calculation of gas concentration for the called channel (device) and subchannel (range). A new linearization curve will be calculated with these x/y-pairs.

Write command

ELIN Kn Mx aaa AAA bbb BBB ... xxx XXX

nth pair of values

2nd pair of values

1st pair of values

Addressed channel n and range x

Code

Response

ELIN 0

Error status

Code

90003752(1) [AK-Commands] 10/98

AK

2 - 91