EXCELL PRECISION CO., LTD.
FM User Manual SME300000119
9
12345
6789
Load cell cable
Shield
Load cell
EXC+
SEN+
SIG+SIG-
EXC-
SEN-
<Chapter 4> Installation

4-1 Load Cell

4-wired (5-wired) Load Cell

Short Pin4&5 to connect with EXC+
Short Pin2&3 to connect with EXC-
Pin1 connects with SIG+
Pin9 connects with SIG-
Pin6, 7, 8 connect with Shield

6-wired (7-wired) Load Cell

Pin5 connects with EXC+
Pin4 connects with SEN+
Pin3 connects with EXC-
Pin2 connects with SEN-
Pin1 connects with SIG+
Pin9 connects with SIG-
Pin6, 7, 8 connect with Shield