2-13
Installation

FIGURE 2-11.Model 444

with

Optional Mounting Bracket.

FIGURE 2-12.Model 444

Dimensional Drawings

Transmittercan be Rotated 90°
Clearance Hole
for ¼-inch Bolt
(eight places)
NOTE
Dimensionsare in inches (millimeters).
Mounting Bracket
¼–20 3½-inch
Bolt (4)
PIPESTAND MOUNTING
5/16–18 U-bolt for
2-inch Pipe (2)
5/16-inch Bolts
(four required,
not furnished) Hole for
5/16-inch
Bolts
(four
places)
PANEL OR SURFACE MOUNTING
2.81 (81)
5.00 (127)
444-1151G, 1151F04A
NOTE
Dimensions are in inches (millimeters).
PermanentTag
(Optional)
4.5 Max.
(114)
4.5 Max.
(114)
Explosion Proof or
IntrinsicSafety Label
(Optional)
Nameplate
Meter Housing
½–14 NPT per
ANSI C80.4 for
Conduitor Sensor
Connection
(two places)
7.5 (191) Max. with Optional
Meter
0.75 (19)
Clearance for
Cover Removal
(Typical)
4.5 Max.
(114)
4.2
(117)
Terminal
Connections
this Side
Terminal
Circuitry
this Side
0.72 (18)
0.36 (9)
0.87 (22)
1.7
(44)
Mounting Holes
¼–20 UNC–2B
0.375 (10) Min. Dp.
(four places)
444-51LTE05A, 51LTG 05A, 51LTF 05A