Rosemount Model 444 Alphaline Temperature Transmitters
6-4
LCD MeterAssembly
Figure 6-3 shows the mounting hardware required to properly install the LCD
meter on a Model 444 transmitter. This mounting hardware may also be used
with the Rosemount universal (analog) meter.
FIGURE 6-3. LCD Meter
Exploded View
LCD Meter
Specifications
Functional
Specifications Input Signal
4–20 mA dc.
Meter Indication
4-digit LCD showing –999 to 9999. A 20-segment bar graph directly represen ts
the 4–20 mA current.
Scaling/Calibration
4 mA Point Limits: –999 to 1000.
Span limits: 0200 to 9999.
The sum of the 4 mA point and span must not exceed 9999.
Adjustments are made using non-interactive zero and span buttons.
Hazardous Locations Certifications
Approved for use with Rosemount Models 444, 751, 1135, 1144,
and 1151.
CoverFoamSpacer
Cover Bushing
Mounting Plate
Meter (Meter may be rotated
in 90 degree increments)
TerminalScrews (Mount
into Transmitter“Test”
TerminalBlock)
MountingScrews
Retaining Straps
MountingScrew into Housing Strap Washer
MountingScrew into
MountingPlate
751-0264B