Fluke 1526 user manual Testing and Replacing the Fuse W, Testing the Fuse

Models: 1526

1 41
Download 41 pages 38.31 Kb
Page 29
Image 29

Fluke 1520

Users Manual

Testing and Replacing the Fuse W

Warning

To avoid electric shock, disconnect the test leads from the inputs before opening the Meter for fuse replacement.

To prevent personal injury or damage to the Meter, install ONLY Fluke specified fuse identified in the “Replacement Parts and Optional Accessories” section.

Before replacing the fuse, test it as d scribed in the next section.

Testing the Fuse

Use the following procedure to test the internal fuse of the Meter.

1.Turn the rotary switch to the Lo eFunction.

2. Connect the test leads to the input terminals and short them together. Press TEST.

3.The display should indicate approximately 0.5 Ω. If the display reads >40 Ω, replace the fuse as described next and

test again.

acf06f.eps

24

Page 29
Image 29
Fluke 1526 user manual Testing and Replacing the Fuse W, Testing the Fuse