Fluke 1526 user manual Fluke

Models: 1526

1 41
Download 41 pages 38.31 Kb
Page 19
Image 19

Fluke 1520

Users Manual

The Meter beeps when the reading is stable. The upper right display shows the same resistance reading as the main display.

When resistance is higher than the maximum display range, the meter reacts as follows:

If the 250V range is selected, the display reads >1000 MΩ.

If the 500V range, is selected, the display reads >2000 MΩ.

If the 1000V range, is selected, the display reads >4000 MΩ.

4.While keeping the probes on the test points, release the TEST button.

The circuit now discharges through the Meter, while the main reading shows the decreasing voltage.

Keep the probes touched to the test points until the circuit is completely discharged (the main display shows - - - -).

The upper right display holds the resistance reading until a new test is started or a different function is selected.

14

Page 19
Image 19
Fluke 1526 user manual Fluke